Measuring Intraocular Pressure in Patients With Keratoconus With and Without Intrastromal Corneal Ring Segments.

J Glaucoma

*Department of Ophthalmology, Hospital Clinico San Carlos ‡Unit of Clinical Methodology and Epidemiology, and Preventive Medicine, Institute of Health Care Research of the Hospital Clinico San Carlos (IdISSC), Faculty of Nursing, Podology and Physiotherapy, University Complutense of Madrid †Cooperative Research Network in Ophthalmology (RETICS), Instituto de Salud Carlos III, Madrid, Spain.

Published: January 2017

Purpose: To compare intraocular pressure (IOP) measurements made using 5 tonometers in keratoconic eyes with and without intrastromal corneal ring segments.

Methods And Patients: This was an observational case series study. A total of 147 eyes of 147 patients with keratoconus, 74 of which had undergone corneal ring segment placement, were prospectively evaluated. IOP was measured using the tonometers Tonopen XL, Pascal dynamic contour tonometer, iCare Pro, ocular response analyzer (ORA), and Goldmann applanation (GAT) in random order. The Bland-Altman method was used to examine interinstrument agreement. Effects on readings of central corneal thickness, corneal curvature, and corneal astigmatism were assessed by multivariate regression analysis.

Results: Smallest mean IOP differences with GAT measurements in eyes without and with ring segments, respectively, were detected for iCare Pro [0.2 (2.9) mm Hg and 0.4 (3.0) mm Hg, P=0.914] and greatest differences for ORA Goldmann-correlated IOP [5.8 (3.3) mm Hg and 6.0 (3.1) mm Hg, P=0.363]. Best agreement with GAT was shown by iCare Pro (ICC=0.829; 95% CI, 0.721-0.896) and worse agreement by ORA corneal-compensated IOP (ICC=-0.145; 95% CI, -0.826 to 0.283). All but the dynamic contour tonometer readings were influenced by central corneal thickness, yet these measurements were affected by the presence of ring segments (P=0.017) and corneal astigmatism (P=0.030). Corneal curvature only affected ORA Goldmann-correlated IOP (P=0.029).

Conclusions: All 5 tonometers provided reliable IOP readings in the keratoconic eyes regardless of the presence of corneal ring segments. iCare Pro readings were most consistent with GAT, whereas ORA readings were least consistent with this reference standard.

Download full-text PDF

Source
http://dx.doi.org/10.1097/IJG.0000000000000549DOI Listing

Publication Analysis

Top Keywords

corneal ring
16
ring segments
16
icare pro
16
corneal
10
intraocular pressure
8
patients keratoconus
8
intrastromal corneal
8
keratoconic eyes
8
dynamic contour
8
contour tonometer
8

Similar Publications

Enhanced Ocular Bioavailability and Prolonged Duration via Hydrophilic Surface Nanocomposite Vesicles for Topical Drug Administration.

Pharmaceutics

November 2024

Guangdong Provincial Key Laboratory of Advanced Drug Delivery, Guangdong Provincial Engineering Center of Topical Precise Drug Delivery System, Guangdong Pharmaceutical University, No. 280 University Town Outer Ring East Road, Guangzhou 510006, China.

Background: Internal ocular diseases, such as macular edema, uveitis, and diabetic macular edema require precise delivery of therapeutic agents to specific regions within the eye. However, the eye's complex anatomical structure and physiological barriers present significant challenges to drug penetration and distribution. Traditional eye drops suffer from low bioavailability primarily due to rapid clearance mechanisms.

View Article and Find Full Text PDF

Acquired Ankyloblepharon Correction Using Ocular Surface and Tarsal Mucous Membrane Grafting in Cicatrizing Ocular Surface Diseases.

Ophthalmic Plast Reconstr Surg

January 2025

Hariram Motumal Nasta & Renu Hariram Nasta Ophthalmic Plastic Surgery Services.

Purpose: To report the long-term surgical outcomes of acquired ankyloblepharon correction using mucous membrane graft.

Methods: Five eyes of 4 patients (median age, 19 years) with acquired ankyloblepharon were managed using eyelid splitting and mucous membrane graft anchored to the recti muscles on the bulbar surface in the respective quadrant and onto the bare tarsal surface. Outcome measures include a change in the palpebral fissure width, ability to fit scleral contact lenses, visual acuity, and cosmesis.

View Article and Find Full Text PDF

Objectives: To demonstrate corneal remodeling after corneal allograft intrastromal ring segment (CAIRS) with an anterior-segment optical coherence tomography (AS-OCT).

Design: A prospective observational single-center study.

Methods: This observational study included keratoconus patients who underwent CAIRS implantation into a stromal tunnel.

View Article and Find Full Text PDF

Artificial intelligence-assisted fitting method using corneal topography outcomes enhances success rate in orthokeratology lens fitting.

Cont Lens Anterior Eye

December 2024

Eye Institute and Department of Ophthalmology, Eye & ENT Hospital, Fudan University, Shanghai 200031, China; NHC Key laboratory of Myopia and Related Eye Diseases, Key Laboratory of Myopia and Related Eye Diseases, Chinese Academy of Medical Sciences, Shanghai, 200031, China; Shanghai Research Center of Ophthalmology and Optometry, Shanghai 200031, China; Shanghai Engineering Research Center of Laser and Autostereoscopic 3D for Vision Care (20DZ2255000), China. Electronic address:

Purpose: Based on ideal outcomes of corneal topography following orthokeratology (OK), an innovative machine learning algorithm for corneal refractive therapy (CRT) was developed to investigate the precision of artificial intelligence (AI)-assisted OK lens fitting.

Methods: A total of 797 eyes that had been fitted with CRT lenses and demonstrated good lens centration with plus power ring intact in their topography were retrospectively included. A comprehensive AI model included spherical refraction, keratometry readings, eccentricity, corneal astigmatism, horizontal visible iris diameter, inferior-superior index, surface asymmetry index, surface regularity index and 8-mm chordal corneal height difference.

View Article and Find Full Text PDF

Purpose: To investigate the effectiveness and safety of customized femtosecond laser-assisted corneal allogenic intrastromal ring segments (CAIRS) implantation in keratoconus patients with a history of intrastromal corneal ring segment (ICRS) explantation.

Setting: Sorocaba Eye Hospital (Hospital Oftalmológico de Sorocaba - HOS), Sorocaba, Brazil.

Design: Prospective case series.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!