Recommendations for Lengthening of Magnetically Controlled Growing Rods in Children With Pacemakers.

J Pediatr Orthop

*Exeter Spine Unit, Princess Elizabeth Orthopaedic Centre, Royal Devon and Exeter NHS Foundation Trust, Exeter ‡The James Cook University Hospital, Middlesbrough, United Kingdom †Faculty of Medicine, University of New South Wales, Sydney, NSW, Australia.

Published: June 2017

Purpose Of The Study: Pacemakers are currently identified as a contraindication for the use of magnetic growth rods (MGRs). This arises from concern that magnetic fields generated by the MGR external remote controller (ERC) during lengthening procedures may induce pacemaker dysfunction. We investigated (1) whether MGR lengthening affects pacemaker function, and (2) if the magnetic field of a pacemaker affects MGR lengthening.

Methods: MGRs were tested in conjunction with an magnetic resonance imaging-compatible pacemaker, which was connected to a virtual patient under continuous cardiac monitoring. To determine whether pacemaker function was affected during MGR lengthening, the electrocardiogram trace was monitored for arrhythmias, whereas an ERC was applied to lengthen the MGRs at varying distances from the pacemaker. To investigate if MGR lengthening was affected by the presence of a pacemaker, at the start and end of the experiment, the ability of the rods to fully elongate and shorten was tested to check for conservation of function.

Results: When the pacemaker was in normal mode, <16 cm away from the activated ERC during MGR lengthening, pacemaker function was affected by the ERC's magnetic forces. At this distance, prophylactically switching the pacemaker to tonic mode before lengthening prevented occurrence of inappropriate pacing discharges. No deleterious effect of the pacemaker's magnetic field on the MGR lengthening mechanism was identified.

Conclusions: Magnetic resonance imaging-compatible pacemakers appear safe for concomitant use with MGRs, provided a pacemaker technician prophylactically switches the pacemaker to tonic function before outpatient lengthening procedures.

Clinical Relevance: This experiment was designed to provide the first safety information on MGR lengthening in children with pacemakers. Although currently a rare clinical scenario, with increasing use of MGRs, this clinical scenario may arise more frequently in the future.

Download full-text PDF

Source
http://dx.doi.org/10.1097/BPO.0000000000000872DOI Listing

Publication Analysis

Top Keywords

mgr lengthening
12
pacemaker
8
pacemaker function
8
mgr
5
recommendations lengthening
4
lengthening magnetically
4
magnetically controlled
4
controlled growing
4
growing rods
4
rods children
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!