AI Article Synopsis

  • * Using bovine bone blocks, researchers tested 4-mm by 10-mm implants at different torques and measured the correlation between these methods.
  • * Results showed a strong inverse correlation between resonance frequency and implant damping (r = -0.98), while the correlation with insertion torque was weak, helping clinicians in determining when to load dental implants.

Article Abstract

Purpose: Various advanced technologies have been designed to estimate dental implant stability, including insertion torque, implant damping using the Periotest, and resonance frequency analysis expressed in implant stability quotient units. This study aimed to establish the relationship between the estimates for these three techniques.

Materials And Methods: Bovine cortical bone blocks were trimmed to square shapes of precisely controlled thickness and density. Thereafter, 4-mm-diameter × 10-mm-long implants were placed in the bone blocks using 30-, 45-, and 60-Ncm insertion torques. Implant damping and resonance frequency analysis were conducted at four angles for each implant. Spearman correlation coefficients were calculated to examine the relationship between insertion torque, resonance frequency analysis, and implant damping measured values.

Results: Resonance frequency analysis and implant damping estimates exhibited a strong inverse correlation (r = -0.98, P < .001). A linear equation was formulated: Periotest value = 15.54 + (-0.26 × implant stability quotient). Implant damping and resonance frequency estimates revealed a weak correlation with insertion torque (P < .05).

Conclusion: This study mathematically defined a linear correlation between implant damping and resonance frequency estimates, thereby providing a guide for clinicians when deciding the loading time for dental implants using the implant damping or resonance frequency estimates.

Download full-text PDF

Source
http://dx.doi.org/10.11607/jomi.4468DOI Listing

Publication Analysis

Top Keywords

resonance frequency
32
implant damping
28
damping resonance
20
frequency analysis
16
implant stability
12
insertion torque
12
frequency estimates
12
implant
11
damping
8
resonance
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!