AI Article Synopsis

  • The study evaluated the "Mapleson D CPAP system" for weaning infants from mechanical ventilation after extended usage, specifically targeting infants who needed ventilation for over 24 hours.
  • Data were collected from 50 infants, analyzing age, weight, and vital signs, with a focus on comparing outcomes pre- and post-use of the CPAP system and during spontaneous ventilation.
  • Results showed that the Mapleson D CPAP system is a safe, effective, and cost-efficient alternative for facilitating the weaning process, despite a notable extubation failure rate of 26%, mainly correlated with infant weight and age.

Article Abstract

Background: Over the last years, we have used a flow-inflating bag circuit with a nasotracheal or nasopharyngeal tube as an interface to deliver effective CPAP support in infants ("Mapleson D CPAP system"). The primary goal of this study was to assess the usefulness of the "Mapleson D CPAP system" for weaning of mechanical ventilation (MV) in infants who received MV over 24 h.

Materials And Methods: All infants who received MV for more than 24 h in the last year were enrolled in the study. Demographic data included age, gender, weight, and admission diagnosis. Heart rate, respiratory rate, blood pressure, and oxygen saturation were measured during MV, 2 h after the nasotracheal Mapleson D CPAP system and 2 h after extubation. Patients were classified into two groups: patients MV more than 48 h, and patients with MV fewer than 48 h. P < 0.05 was considered statistically significant.

Results: A total of 50 children were enrolled in the study, with a median age was 34 ± 45 months (range, 1-59 months) and median weight was 11.98 ± 9.31 kg (range, 1-48 kg). Median duration of MV was 480 h (range, 2-570). There were no significant differences in PaO2, PaCO2, and pH among MV, 2 h after the nasotracheal Mapleson D CPAP system and 2 h after extubation and spontaneous ventilation with the nasopharyngeal Mapleson D CPAP system or with nasal prongs. The overall extubation failure rate was 26% (n = 13). Weight and age were significantly associated with extubation failure (P < 0.05).

Conclusions: The Mapleson D CPAP system, in our opinion, is a useful and safe alternative to more complex and expensive noninvasive CPAP and BiPAP weaning from MV in infants.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC5006332PMC
http://dx.doi.org/10.4103/0970-2113.188972DOI Listing

Publication Analysis

Top Keywords

mapleson cpap
20
cpap system
20
cpap
9
weaning mechanical
8
mechanical ventilation
8
"mapleson cpap
8
cpap system"
8
infants received
8
enrolled study
8
nasotracheal mapleson
8

Similar Publications

Article Synopsis
  • The study evaluated the "Mapleson D CPAP system" for weaning infants from mechanical ventilation after extended usage, specifically targeting infants who needed ventilation for over 24 hours.
  • Data were collected from 50 infants, analyzing age, weight, and vital signs, with a focus on comparing outcomes pre- and post-use of the CPAP system and during spontaneous ventilation.
  • Results showed that the Mapleson D CPAP system is a safe, effective, and cost-efficient alternative for facilitating the weaning process, despite a notable extubation failure rate of 26%, mainly correlated with infant weight and age.
View Article and Find Full Text PDF

Objectives: Ventilation of patients undergoing procedures in the prone position represents a challenge for the anesthesiologist, especially when trying to avoid tracheal intubation. This study aimed to test the effectiveness and safety of a prototype designed for pediatric facial mask ventilation in the prone position.

Material And Methods: A prospective descriptive study was conducted on 105 sedations performed in 3 children scheduled for radiotherapy treatment of posterior fossa desmoplastic medulloblastoma (6 and 4 y.

View Article and Find Full Text PDF

Background And Aims: Induction of general anaesthesia per se as also the use of 100% oxygen during induction of anaesthesia, results in the development of atelectasis in dependent lung regions within minutes of anaesthetic induction. We aimed to assess the effect of application of a continuous positive airway pressure (CPAP) of 5 cm H2O during pre-oxygenation and induction of anaesthesia on the period of apnoea before the occurrence of clinically significant desaturation.

Methods: In this prospective, randomised, and double-blind study, 40 patients posted for elective surgery were enrolled.

View Article and Find Full Text PDF

New airway device for ventilation and monitoring in pediatric patients undergoing MRI study.

J Clin Monit Comput

February 2012

Department of Anesthesiology, Cincinnati Children's Hospital Medical Center, 3333 Burnet Avenue, ML 2001, Cincinnati, OH 45209, USA.

A method of administering continuous positive airway pressure via a new airway device to prevent upper airway obstruction and preserve spontaneous respiration under total intravenous anesthesia has been adapted for children undergoing deep sedation for MRI studies. Presented herein is a retrospective study of 45 pediatric patients, ages 5 months to 7 years, who underwent an MRI study under general anesthesia using a modified nasal vestibule airway (NVA®), a pressure-sealing nasal cannula that can be used in conjunction with an anesthesia circuit to deliver nasal-CPAP during anesthesia. After inhalation induction of anesthesia with sevoflurane, an intravenous infusion of propofol was used to maintain anesthesia.

View Article and Find Full Text PDF

Introduction: This case report describes the anaesthetic management of exploratory thoracoscopy and alternating one lung ventilation (OLV) in a dog with a pulmonary bulla, and the application of continuous positive airway pressure (CPAP) to the non-ventilated lung for preventing and treating hypoxia.

Case History: A 6-year-old, male castrated Border collie was scheduled for exploratory thoracoscopy to investigate spontaneous pnemothorax that had not resolved with repeated suction. Specific requirements for the thoracoscopy were alternating OLV to allow the surgical access to the right middle lobe and its removal, and the examination of the left hemithorax to rule out the presence of other lesions.

View Article and Find Full Text PDF