A young man with a rare unstable haemoglobinopathy presented with a high fever, worsening shortness of breath and abdominal pain. At triage his pulse oximetry (SpO2) suggested that his blood oxygen saturation was 84% at room air. However, an arterial blood gas (ABG) oxygen saturation reading (SaO2) was 100%. The significant disparity between the two measurements demonstrates that using pulse oximetry in some unstable haemoglobinopathies may significantly underestimate the actual reading. This error is most probably due to the structural differences in the variant haemoglobin causing light to be absorbed at a different wavelength beyond the normal range of the oximeter. Haemoglobinopathies affect about 7% of the world's population and is often asymptomatic; so, there may be many more undiagnosed cases. Therefore, clinicians may confirm low SpO2 readings with an ABG and, where there is significant disparity with no obvious extrinsic cause, they should consider haemoglobinopathies.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC5020765PMC
http://dx.doi.org/10.1136/bcr-2016-216787DOI Listing

Publication Analysis

Top Keywords

pulse oximetry
12
oxygen saturation
8
oximetry error
4
error patient
4
patient santa
4
santa ana
4
ana haemoglobinopathy
4
haemoglobinopathy young
4
young man
4
man rare
4

Similar Publications

Background: Evidence supports the common incidence of sleep disturbance in opioid use disorder (OUD) as a potential marker of disrupted orexin system functioning. This study evaluated the initial safety and tolerability of a challenge dose of lemborexant, a dual orexin antagonist, as an adjunct to buprenorphine/naloxone.

Methods: Patients (18-65 years old) with OUD receiving sublingual buprenorphine/naloxone, with a Pittsburgh Sleep Quality Index total score of 6 or higher, were recruited from outpatient clinics.

View Article and Find Full Text PDF

Background: Cerebral autoregulation is a robust regulatory mechanism that stabilizes cerebral blood flow in response to reduced blood pressure, thereby preventing cerebral ischaemia. Scientists have long believed that cerebral autoregulation also stabilizes cerebral blood flow against increases in intracranial pressure, which is another component that determines cerebral perfusion pressure. However, this idea was inconsistent with the complex pathogenesis of normal pressure hydrocephalus, which includes components of chronic cerebral ischaemia due to mild increases in intracranial pressure.

View Article and Find Full Text PDF

: Attaining adequate oxygenation in critically ill patients undergoing invasive ventilation necessitates intense monitoring through pulse oximetry (SpO) and frequent manual adjustments of ventilator settings like the fraction of inspired oxygen (FiO) and the level of positive end-expiratory pressure (PEEP). Our aim was to compare the quality of oxygenation with the use of automated ventilation provided by INTELLiVENT-Adaptive Support Ventilation (ASV) vs. ventilation that is not automated, i.

View Article and Find Full Text PDF

Validation of Aerobic Capacity (VO2max) and Pulse Oximetry in Wearable Technology.

Sensors (Basel)

January 2025

School of Integrated Health Sciences, Department of Kinesiology and Nutrition Sciences, University of Nevada, Las Vegas, NV 89154, USA.

Introduction: As wearable technology becomes increasingly popular and sophisticated, independent validation is needed to determine its accuracy and potential applications. Therefore, the purpose of this study was to evaluate the accuracy (validity) of VO2max estimates and blood oxygen saturation measured via pulse oximetry using the Garmin fēnix 6 with a general population participant pool.

Methods: We recruited apparently healthy individuals (both active and sedentary) for VO2max (n = 19) and pulse oximetry testing (n = 22).

View Article and Find Full Text PDF

Accuracy of Rhythm Diagnostic Systems' MultiSense in Detection of Arterial Oxygen Saturation and Respiratory Rate During Hypoxia in Humans: Effects of Skin Color and Device Localization.

Sensors (Basel)

December 2024

Biomedicine Research Center of Strasbourg (CRBS), UR 3072, "Mitochondria, Oxidative Stress and Muscle Plasticity", Faculty of Medicine, University of Strasbourg, 67000 Strasbourg, France.

The continuous monitoring of oxygen saturation (SpO) and respiratory rates (RRs) are major clinical issues in many cardio-respiratory diseases and have been of tremendous importance during the COVID-19 pandemic. The early detection of hypoxemia was crucial since it precedes significant complications, and SpO follow-up allowed early hospital discharge in patients needing oxygen therapy. Nevertheless, fingertip devices showed some practical limitations.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!