Introduction: Recently, a new urodynamic method for the assessment of stress urinary incontinence called high definition urethral pressure profilometry (HD-UPP) has been introduced. This method combines a novel microtip catheter with advanced signal processing to enable spatial data location and the reconstruction of a pressure image inside the urethra. In order to assess the reproducibility of HD-UPP data, we statistically evaluate HD-UPP datasets and compare them to data from a double balloon air-charged system.
Materials And Methods: Both catheters are used on sedated female minipigs. Data from the microtip catheter are processed through a signal reconstruction algorithm, urodynamic features are extracted, and compared to the air-charged system. Reproducibility of HD-UPP data is assessed by statistically evaluating consecutive, intra-individual datasets.
Results: HD-UPP delivers results in agreement with previous comparisons of microtip and air-charged systems. The average deviation of two consecutive, intra-individual pressure images is very low at 7 cm H O.
Conclusions: HD-UPP provides physicians with detailed information on the pressure distribution inside the urethra. Through comparison with an air-charged catheter, it is shown that HD-UPP delivers results in agreement with previous studies on the comparison of microtip and air-charged catheters. It provides excellent reproducibility, as the difference between sequentially measured profiles from the same minipig is significantly lower than the one between profiles from different minipigs.