Purpose: To evaluate the efficacy of infrared (IR) ocular thermography in screening for dry eye disease (DED).

Methods: IR ocular thermography was performed on 62 dry eye and 63 age- and sex-matched control subjects. Marking of ocular surface and temperature acquisition was done using a novel 'diamond' demarcation method. 30 static- and 30 dynamic-metrics were studied and receiver operating characteristic curves were plotted. Efficacy of the temperature metrics in detecting DED were evaluated singly and in combination in terms of their area under the curve (AUC), Youden's index and discrimination power (DP).

Results: Absolute temperature of the extreme nasal conjunctiva 5s and 10s after eye opening were best detectors for DED. With threshold value for the first metric set at 34.7°C, sensitivity and specificity was 87.1% (95% CI: 76.2-94.3%) and 50.8% (95% CI: 37.9-63.6%) respectively. With threshold value for the second metric set at 34.5°C, sensitivity and specificity was 77.6% (95% CI: 64.7-87.5%) and 61.9% (95% CI: 48.8-73.9%) respectively. The two metrics had moderate accuracy and limited performances with AUC of 72% (95% CI: 63-81%) and 73% (95% CI: 64-82%); Youden index of about 0.4 and DP of 1.07 and 1.05 respectively. None of the dynamic metrics was good detector for DED. Combining metrics was not able to increase the AUC.

Conclusions: This work suggests some utility for the application of IR ocular thermography for evaluation of dry eye patients.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.clae.2016.08.004DOI Listing

Publication Analysis

Top Keywords

dry eye
16
ocular thermography
16
screening dry
8
eye disease
8
infrared ocular
8
metric set
8
sensitivity specificity
8
95%
6
eye
5
ocular
5

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!