Background: The American National Standards Institute (ANSI) provides standards used to govern standardization of all hearing aids. If hearing aids do not meet specifications, there are potential negative implications for hearing aid users, professionals, and the industry. Recent literature has not investigated the proportion of new hearing aids in compliance with the ANSI specifications for quality control standards when they arrive in the clinic before dispensing.
Purpose: The aims of this study were to determine the percentage of new hearing aids compliant with the relevant ANSI standard and to report trends in electroacoustic analysis data.
Research Design: New hearing aids were evaluated for quality control via the ANSI S3.22-2009 standard. In addition, quality control of directional processing was also assessed.
Study Sample: Seventy-three behind-the-ear hearing aids from four major manufacturers, that were purchased for clinical patients were evaluated before dispensing.
Data Collection And Analysis: Audioscan Verifit (version 3.1) hearing instrument fitting system was used to complete electroacoustic analysis and directional processing evaluation of the hearing aids. Frye's Fonix 8000 test box system (Fonix 8000) was also used to cross-check equivalent input noise (EIN) measurements. These measurements were then analyzed for trends across brands and specifications.
Results: All of the hearing aids evaluated were found to be out of specification for at least one measure. EIN and attack and release times were the measures most frequently out of specification. EIN was found to be affected by test box isolation for two of the four brands tested. Systematic discrepancies accounted for ∼93% of the noncompliance issues, while unsystematic quality control issues accounted for the remaining 7%.
Conclusions: The high number of systematic discrepancies between the data collected and the specifications published by the manufacturers suggests there are clear issues related to the specific protocols used for quality control testing. These issues present a significant barrier for hearing aid dispensers when attempting to accurately determine if a hearing aid is functioning appropriately. The significant number of unsystematic discrepancies supports the continued importance of quality control measures of new and repaired hearing aids to ensure that the device is functioning properly before it is dispensed and to avoid future negative implications of fitting a faulty device.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.3766/jaaa.15027 | DOI Listing |
JAMA Netw Open
January 2025
Division of General Internal Medicine and Primary Care, Brigham and Women's Hospital, Boston, Massachusetts.
Importance: Nearly all Medicare Advantage (MA) plans offer dental, vision, and hearing benefits not covered by traditional Medicare (TM). However, little is known about MA enrollees' use of those benefits or how much they cost MA insurers or enrollees.
Objective: To estimate use, out-of-pocket (OOP) spending, and insurer payments for dental, hearing, and vision services among Medicare beneficiaries.
Heliyon
January 2025
Department of Laboratory Medicine, Fujian Medical University, Fuzhou, 350122, China.
Presbycusis, also referred to as age-related hearing loss, poses a substantial burden on both individuals and society. The hallmark of presbycusis is a progressive decrease in auditory sensitivity. Irreversible hearing loss occurs due to the limited regenerative capacity of spiral neurons and peripheral cochlear hair cells (HCs).
View Article and Find Full Text PDFJ Korean Med Sci
January 2025
Department of Otolaryngology-Head and Neck Surgery, College of Medicine, Hanyang University, Seoul, Korea.
Background: Hearing level reference values based on the results of recent audiometry have not been established for the general population of South Korea. This study aimed to evaluate the mean hearing levels of each age group and to measure the annual progression of hearing loss.
Methods: We used the database of the eighth and ninth Korea National Health and Nutrition Examination Survey from 2020 to 2022, and included participants with normal tympanic membranes and without occupational noise exposure.
BMJ Open
December 2024
National Institute of Public Health, University of Southern Denmark, Copenhagen, Denmark.
Introduction: Individuals with hearing loss and hearing aid users report higher levels of listening effort and fatigue in daily life compared with those with normal hearing. However, there is a lack of objective measures to evaluate these experiences in real-world settings. Recent studies have found that higher sound pressure levels (SPL) and lower signal-to-noise ratios (SNR) are linked to increased heart rate and decreased heart rate variability, reflecting the greater effort required to process auditory information.
View Article and Find Full Text PDFJ Clin Med
December 2024
Unit of Otorhinolaryngology, Department of Adult and Development Age Human Pathology "Gaetano Barresi", University of Messina, 98122 Messina, Italy.
This study aims to describe the stress levels experienced by parents of children with hearing loss who use conventional hearing aids or cochlear implants, and to assess the correlation between parental stress and the auditory skills acquired by the children. The study was conducted at the Policlinic "Gaetano Martino" in Messina, evaluating data from 42 pairs of parents of children using hearing aids or cochlear implants. Parents completed the LittlEARS Auditory Questionnaire (LEAQ) and the Parental Stress Scale (PSS) 18 months after the initial device (hearing aid or cochlear implant) had been activated.
View Article and Find Full Text PDFEnter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!