Background: Radiofrequency (RF) technology has improved detection of retained surgical sponges with a reported 100% sensitivity and specificity. However, the potential for interactions of the RF signals emitted by the detection system with cardiac implantable electronic devices (CIEDs) or temporary pacemakers may limit its use in those patients with these devices. This study investigated whether RF detection technology causes interference or clinically significant changes in the programmed settings of implanted pacemakers and defibrillators or temporary epicardial pacemakers.

Methods: Fifty patients who were scheduled either for CIED removal or placement of a temporary epicardial pacemaker (at the time of open heart surgery) were recruited for this study. Device settings and measurements from separate interrogations before and after scanning with the RF detection system were compared. For the temporary pacemakers, we observed for any changes in hemodynamics or signs of pacing interference.

Results: Twenty (40%) pacemakers, 20 (40%) implantable cardioverter defibrillators, and 10 (20%) temporary pacemakers were analyzed in this study. During scanning, no signal interference was detected in any permanent device, and there were no significant changes in programmed settings after scanning with the RF detection system. However, pacing inhibition was detected with temporary pacing systems when programmed to a synchronous mode (DDD).

Conclusions: RF detection technology can be safely used to scan for retained surgical sponges in patients with permanent CIEDs and temporary pacemakers set to asynchronous mode.

Download full-text PDF

Source
http://dx.doi.org/10.1111/pace.12938DOI Listing

Publication Analysis

Top Keywords

temporary pacemakers
20
detection technology
12
detection system
12
cardiac implantable
8
implantable electronic
8
electronic devices
8
temporary
8
retained surgical
8
surgical sponges
8
cieds temporary
8

Similar Publications

Background: Immune checkpoint inhibitors (ICIs) are effective antineoplastic agents but can cause adverse effects in many organ systems. Cardiovascular toxicities include arrhythmias, myocarditis, heart failure, takotsubo syndrome, pericarditis, coronary artery disease, and vasculitis.

Case Summary: A 66-year-old woman with Stage 3C2 endometrial carcinoma presented for her second cycle of pembrolizumab, carboplatin, and paclitaxel.

View Article and Find Full Text PDF

The increasing use of immune checkpoint inhibitors in clinical practice is associated with a broad spectrum of immune-related adverse events, such as cardiac, rheumatological and neurological toxicities. Myocarditis is a life-threatening complication, and the concurrent development of myocarditis, myositis and/or myasthenia leads to difficulties in diagnosis, management and treatment. We describe a case presenting with this triple M overlap syndrome following pembrolizumab treatment.

View Article and Find Full Text PDF

Validity and Accuracy of the Derived Left Ventricular End-Diastolic Pressure in Impella 5.5.

Circ Heart Fail

January 2025

Department of Cardiothoracic, Transplantation and Vascular Surgery, Hannover Medical School, Germany. (R.P., J.S.H., D.B., A.S.M., M.H., A.Z., G.D., J.D.S., A.F.P., A.W., A.R., B.S.).

Background: Consensus regarding on-support evaluation and weaning concepts from Impella 5.5 support is scarce. The derived left ventricular end-diastolic pressure (dLVEDP), estimated by device algorithms, is a rarely reported tool for monitoring the weaning process.

View Article and Find Full Text PDF

Introduction: Haemodynamic atrioventricular delay (AVD) optimisation has primarily focussed on signals that are not easy to acquire from a pacing system itself, such as invasive left ventricular catheterisation or arterial blood pressure (ABP). In this study, standard clinical central venous pressure (CVP) signals are tested as a potential alternative.

Methods: Sixteen patients with a temporary pacemaker after cardiac surgery were studied.

View Article and Find Full Text PDF

Background: Atrial infarction is a complication of myocardial infarction with ventricular infarction; however, isolated atrial infarction (IAI) has rarely been reported. Herein, we report a case of IAI associated with sick sinus syndrome and atrial fibrillation (AF).

Case Summary: An 83-year-old woman was brought to the emergency department with a complaint of general malaise.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!