A comparison of facemask and respirator filtration test methods.

J Occup Environ Hyg

b Nelson Laboratory , Salt Lake City, Salt Lake City , Utah.

Published: February 2017

NIOSH published a Federal Register Notice to explore the possibility of incorporating FDA required filtration tests for surgical masks (SMs) in the 42 CFR Part 84 respirator certification process. There have been no published studies comparing the filtration efficiency test methods used for NIOSH certification of N95 filtering facepiece respirators (N95 FFRs) with those used by the FDA for clearance of SMs. To address this issue, filtration efficiencies of "N95 FFRs" including six N95 FFR models and three surgical N95 FFR models, and three SM models were measured using the NIOSH NaCl aerosol test method, and FDA required particulate filtration efficiency (PFE) and bacterial filtration efficiency (BFE) methods, and viral filtration efficiency (VFE) method. Five samples of each model were tested using each method. Both PFE and BFE tests were done using unneutralized particles as per FDA guidance document. PFE was measured using 0.1 µm size polystyrene latex particles and BFE with ∼3.0 µm size particles containing Staphylococcus aureus bacteria. VFE was obtained using ∼3.0 µm size particles containing phiX 174 as the challenge virus and Escherichia coli as the host. Results showed that the efficiencies measured by the NIOSH NaCl method for "N95 FFRs" were from 98.15-99.68% compared to 99.74-99.99% for PFE, 99.62-99.9% for BFE, and 99.8-99.9% for VFE methods. Efficiencies by the NIOSH NaCl method were significantly (p = <0.05) lower than the other methods. SMs showed lower efficiencies (54.72-88.40%) than "N95 FFRs" measured by the NIOSH NaCl method, while PFE, BFE, and VFE methods produced no significant difference. The above results show that the NIOSH NaCl method is relatively conservative and is able to identify poorly performing filtration devices. The higher efficiencies obtained using PFE, BFE and VFE methods show that adding these supplemental particle penetration methods will not improve respirator certification.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC7157953PMC
http://dx.doi.org/10.1080/15459624.2016.1225157DOI Listing

Publication Analysis

Top Keywords

filtration efficiency
16
niosh nacl
12
test methods
8
methods niosh
8
fda required
8
"n95 ffrs"
8
n95 ffr
8
ffr models
8
models three
8
measured niosh
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!