Purpose: The subretinal Alpha IMS visual implant is a CE-approved medical device for restoration of visual functions in blind patients with end-stage outer retina degeneration. We present a method to test the function of the implant objectively in vivo using standard electroretinographic equipment and to assess the devices' parameter range for an optimal perception.

Methods: Subretinal implant Alpha IMS (Retina Implant AG, Reutlingen, Germany) consists of 1500 photodiode-amplifier-electrode units and is implanted surgically into the subretinal space in blind retinitis pigmentosa patients. The voltages that regulate the amplifiers' sensitivity (V ) and gain (V ), related to the perception of contrast and brightness, respectively, are adjusted manually on a handheld power supply device. Corneally recorded implant responses (CRIR) to full-field illumination with long duration flashes in various implant settings for brightness gain (V ) and amplifiers' sensitivity (V ) are measured using electroretinographic setup with a Ganzfeld bowl in a protocol of increasing stimulus luminances up to 1000 cd/m.

Results: CRIRs are a meaningful tool for assessing the transfer characteristic curves of the electronic implant in vivo monitoring the implants' voltage output as a function of log luminance in a sigmoidal shape. Changing the amplifiers' sensitivity (V ) shifts the curve left or right along the log luminance axis. Adjustment of the gain (V ) changes the maximal output. Contrast perception is only possible within the luminance range of the increasing slope of the function.

Conclusions: The technical function of subretinal visual implants can be measured objectively using a standard electroretinographic setup. CRIRs help the patient to optimise the perception by adjusting the gain and luminance range of the device and are a useful tool for clinicians to objectively assess the function of subretinal visual implants in vivo.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC5052310PMC
http://dx.doi.org/10.1007/s10633-016-9557-7DOI Listing

Publication Analysis

Top Keywords

subretinal visual
12
visual implants
12
amplifiers' sensitivity
12
corneally recorded
8
implant
8
recorded implant
8
implant responses
8
alpha ims
8
standard electroretinographic
8
electroretinographic setup
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!