Fatal cobalt toxicity after total hip arthroplasty revision for fractured ceramic components.

Clin Toxicol (Phila)

b Division of Medical Toxicology, Department of Emergency Medicine , University of Pittsburgh, Pittsburgh , PA , USA.

Published: November 2016

Context: Post-arthroplasty metallosis, which refers to metallic corrosion and deposition of metallic debris in the periprosthetic soft tissues of the body, is an uncommon complication. Systemic cobalt toxicity post-arthroplasty is extremely rare. The few known fatal cases of cobalt toxicity appear to be a result of replacing shattered ceramic heads with metal-on-metal or metal-on-polyethylene implants. Friction between residual shards of ceramic and cobalt-chromium implants allows release of cobalt into the synovial fluid and bloodstream, resulting in elevated whole blood cobalt levels and potential toxicity.

Case Details: This is a single patient chart review of a 60-year-old woman with prior ceramic-on-ceramic right total hip arthroplasty complicated by fractured ceramic components and metallosis of the joint. She underwent synovectomy and revision to a metal-on-polyethylene articulation. Ten months post-revision, she presented to the emergency department (ED) with right hip pain, dyspnea, worsening hearing loss, metallic dysgeusia, and weight loss. Chest CTA revealed bilateral pulmonary emboli (PE), and echocardiogram revealed new cardiomyopathy with global left ventricular hypokinesis with an ejection fraction (EF) of 35-40% inconsistent with heart strain from PE. Whole blood cobalt level obtained two days into her admission was 424.3 mcg/L and 24-h urine cobalt level was 4830.5 mcg/L. Although the patient initially clinically improved with regard to her PE and was discharged to home on hospital day 5, she returned 10 days later with a right hip dislocation and underwent closed reduction of the hip. The patient subsequently decompensated, developing cardiogenic shock, and respiratory failure. She went into pulseless electrical activity (PEA) and expired. Autopsy revealed an extensive metallic effusion surrounding the right hip prosthesis that tested positive for cobalt (41,000 mcg/L). There was also cobalt in the heart muscle tissue (2.5 mcg/g). A whole blood cobalt level obtained two days before she expired was 641.6 mcg/L.

Discussion: This is a case of fatal cobalt-induced cardiomyopathy in a patient whose ceramic components of a total hip arthroplasty fractured causing metallosis with worsening cobalt toxicity. We recommend that when a fractured device is revised with a prosthesis with cobalt-chromium components, whole blood and urine cobalt measurements should be obtained and periodically monitored to evaluate for rising concentrations. Providers should be aware of clinical signs and symptoms of cobalt toxicity in patients who have prostheses with cobalt-chromium components. If suspected, toxicology and orthopedics should be involved for possible chelation and removal of the prosthesis.

Download full-text PDF

Source
http://dx.doi.org/10.1080/15563650.2016.1214274DOI Listing

Publication Analysis

Top Keywords

cobalt toxicity
20
total hip
12
hip arthroplasty
12
ceramic components
12
cobalt
12
blood cobalt
12
cobalt level
12
fractured ceramic
8
level days
8
urine cobalt
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!