Delivered dose uncertainty analysis at the tumor apex for ocular brachytherapy.

Med Phys

Department of Medical Physics, Cross Cancer Institute, Edmonton, Alberta T6G 1Z2, Canada and Department of Oncology, University of Alberta, Edmonton, Alberta T6G 2R3, Canada.

Published: August 2016

Purpose: To estimate the total dosimetric uncertainty at the tumor apex for ocular brachytherapy treatments delivered using 16 mm Collaborative Ocular Melanoma Study (COMS) and Super9 plaques loaded with (125)I seeds in order to determine the size of the apex margin that would be required to ensure adequate dosimetric coverage of the tumor.

Methods: The total dosimetric uncertainty was assessed for three reference tumor heights: 3, 5, and 10 mm, using the Guide to the expression of Uncertainty in Measurement/National Institute of Standards and Technology approach. Uncertainties pertaining to seed construction, source strength, plaque assembly, treatment planning calculations, tumor height measurement, plaque placement, and plaque tilt for a simple dome-shaped tumor were investigated and quantified to estimate the total dosimetric uncertainty at the tumor apex. Uncertainties in seed construction were determined using EBT3 Gafchromic film measurements around single seeds, plaque assembly uncertainties were determined using high resolution microCT scanning of loaded plaques to measure seed positions in the plaques, and all other uncertainties were determined from the previously published studies and recommended values. All dose calculations were performed using plaque simulator v5.7.6 ophthalmic treatment planning system with the inclusion of plaque heterogeneity corrections.

Results: The total dosimetric uncertainties at 3, 5, and 10 mm tumor heights for the 16 mm COMS plaque were 17.3%, 16.1%, and 14.2%, respectively, and for the Super9 plaque were 18.2%, 14.4%, and 13.1%, respectively (all values with coverage factor k = 2). The apex margins at 3, 5, and 10 mm tumor heights required to adequately account for these uncertainties were 1.3, 1.3, and 1.4 mm, respectively, for the 16 mm COMS plaque, and 1.8, 1.4, and 1.2 mm, respectively, for the Super9 plaque. These uncertainties and associated margins are dependent on the dose gradient at the given prescription depth, thus resulting in the changing uncertainties and margins with depth.

Conclusions: The margins determined in this work can be used as a guide for determining an appropriate apex margin for a given treatment, which can be chosen based on the tumor height. The required margin may need to be increased for more complex scenarios (mushroom shaped tumors, tumors close to the optic nerve, oblique muscle related tilt, etc.) than the simple dome-shaped tumor examined and should be chosen on a case-by-case basis. The sources of uncertainty contributing most significantly to the total dosimetric uncertainty are seed placement within the plaques, treatment planning calculations, tumor height measurement, and plaque tilt. This work presents an uncertainty-based, rational approach to estimating an appropriate apex margin.

Download full-text PDF

Source
http://dx.doi.org/10.1118/1.4959540DOI Listing

Publication Analysis

Top Keywords

total dosimetric
20
dosimetric uncertainty
16
tumor apex
12
apex margin
12
tumor heights
12
treatment planning
12
tumor height
12
tumor
11
plaque
11
apex ocular
8

Similar Publications

Article Synopsis
  • The study examined the risk of radiation esophagitis (RE) in non-small cell lung cancer (NSCLC) patients undergoing immunochemotherapy followed by radiotherapy (ICI-RT) and sought to create a predictive nomogram for RE risk.
  • Researchers analyzed data from 102 patients, identifying key factors (like maximum and mean radiation doses to the esophagus) through logistic regression that significantly correlated with the development of RE.
  • The nomogram showed strong predictive ability in both training (AUC: 0.918) and validation (AUC: 0.833) groups, indicating its potential as a useful tool in clinical practice for predicting RE in this patient population.
View Article and Find Full Text PDF

Background/aim: HyperArc (HA) is an automated planning technique enabling single-isocenter brain stereotactic radiotherapy (SRT); however, dosimetric outcomes may be influenced by the planner's expertise. This study aimed to assess the impact of institutional experience on the plan quality of HA-SRT for both single and multiple brain metastases.

Materials And Methods: Twenty patients who underwent HA-SRT for single metastasis between 2020 and 2021 comprised the earlier group, while those treated between 2022 and 2024 constituted the later group.

View Article and Find Full Text PDF

Development of a prediction model for tube feeding dependence in HPV-associated oropharyngeal cancer patients undergoing chemoradiotherapy.

Oral Oncol

December 2024

Radiation Medicine Program, Princess Margaret Cancer Centre, M5G 2M9, Toronto, Ontario, Canada; Department of Radiation Oncology, Princess Margaret Cancer Centre, University of Toronto, M5G 2M9 Toronto, Ontario. Electronic address:

Objectives: This study aimed to develop a prediction model for feeding tube dependence in a large homogenous cohort of HPV-associated oropharyngeal squamous cell carcinoma (HPV + OPSCC) patients receiving chemoradiotherapy (CRT). We further aimed to externally validate three previously published feeding tube prediction models on this cohort.

Materials And Methods: p16-confirmed HPV + OPSCC patients treated with definitive CRT at a tertiary cancer centre between April 2017 and February 2022 were identified.

View Article and Find Full Text PDF

. Adult medulloblastoma (AMB) patients should receive postoperative craniospinal irradiation (CSI) as a standard treatment. Volumetric intensity-modulated arc therapy (VMAT) is a promising method for CSI.

View Article and Find Full Text PDF

Objectives: Spatially fractionated radiation therapy (SFRT) intentionally delivers a heterogeneous dose distribution characterized by alternating regions of high and low doses throughout a tumor. This modality may enhance response to subsequent whole tumor radiation in bulky and radioresistant lesions that are historically less responsive to conventional radiation doses alone. The current study presents a single institution experience with modern era SFRT using predominantly a volumetric modulated arc therapy (VMAT) lattice technique.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!