Aims: A long-detection interval (LDI) (30/40 intervals) has been proved to be superior to a standard-detection interval (SDI) (18/24 intervals) in terms of reducing unnecessary implantable cardioverter defibrillator (ICD) therapies. To better evaluate the different impact of LDI and anti-tachycardia pacing (ATP) on reducing painful shocks, we assessed all treated episodes in the ADVANCE III trial.

Methods And Results: A total of 452 fast (200 ms < cycle length ≤ 320 ms) arrhythmic episodes were recorded: 284 in 138 patients in the SDI arm and 168 in 82 patients in the LDI arm (106/452 inappropriate detections). A total of 346 fast ventricular tachycardias (FVT) were detected in 169 patients: 208 in 105 patients with SDI and 138 in 64 patients with LDI. Setting LDI determined a significant reduction in appropriate but unnecessary therapies [208 in SDI vs. 138 in LDI; incidence rate ratio (IRR): 0.61 (95% CI 0.45-0.83), P = 0.002]. Anti-tachycardia pacing determined another 52% reduction in unnecessary shocks [208 in SDI with hypothetical shock-only programming vs. 66 in LDI with ATP; IRR: 0.37 (95% CI 0.25-0.53, P < 0.001)]. The efficacy of ATP in terminating FVT was 63% in SDI and 52% in LDI (P = 0.022). No difference in the safety profile (acceleration/degeneration and death/cardiovascular hospitalizations) was observed between the two groups.

Conclusion: The combination of LDI and ATP during charging is extremely effective and significantly reduces appropriate but unnecessary therapies. The use of LDI alone yielded a 39% reduction in appropriate but unnecessary therapies; ATP on top of LDI determined another 52% reduction in unnecessary shocks. The strategy of associating ATP and LDI could be considered in the majority of ICD recipients.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC5408998PMC
http://dx.doi.org/10.1093/europace/euw032DOI Listing

Publication Analysis

Top Keywords

anti-tachycardia pacing
12
unnecessary shocks
12
ldi
12
appropriate unnecessary
12
unnecessary therapies
12
long-detection interval
8
reducing unnecessary
8
advance iii
8
138 patients
8
patients sdi
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!