Preparing Drugs for Infusion Via Syringe Pump: A Key Step to Ensure Homogeneous Concentration.

Crit Care Nurse

Bruno Garrigue is an anesthesia nurse and head of the prehospital emergency medicine service at Centre Hospitalier Sud-Francilien, Corbeil-Essonnes, France.Yann Dehu is an anesthesia nurse working in a surgical unit and a clinical research associate at Centre Hospitalier Sud-Francilien.Fabrice Girault is a registered nurse working in a prehospital emergency medicine service, Centre Hospitalier Sud-Francilien.Bruno Figadère is a medicinal chemist at Centre Nationale de Recherches Scientifiques (CNRS)/Universite Paris Sud-Faculte de Pharmacie Joint Laboratory, Chatenay-Malabry, France.Karine Leblanc is a chemical engineer at CNRS/Univ Paris Sud-Faculte de Pharmacie Joint Laboratory.Nicolas Briole is an emergency medicine physician at Centre Hospitalier Sud-Francilien.Georges Antoine Capitani is head of the acute care department at Centre Hospitalier Sud-Francilien.Steven Lagadec is a nurse in the prehospital emergency service and a clinical research associate at Centre Hospitalier Sud-Francilien.François-Xavier Laborne is a prehospital emergency practitioner and a methodologist and biostatistician in the clinical research unit of the Centre Hospitalier Sud-Francilien.

Published: August 2016

Objective: Preparation of drug solutions used with electronic syringe infusion pumps plays a crucial role in the delivery of an accurate drug concentration. Is there a correlation between drug concentrations during syringe pump infusion and preparation protocols?

Method: Norepinephrine, insulin, and sufentanil were prepared in 3 different ways: (1) the drug was taken from the vial, then the solvent was added followed by an air bubble, and mixing was performed by turning the syringe top-to-bottom in a 180° shaking movement 5 consecutive times; (2) the drug was taken from the vial, then the solvent was added and not mixed; and (3) the solvent was taken from a stock solution, then the drug was added and not mixed. Concentrations of drugs were determined at different times during administration by reverse-phase high-performance liquid chromatography with ultraviolet detection. All analyses were performed in triplicate and were based on measurement of peak areas.

Results: With no shaking of the syringe, the concentration of the injected drugs varies widely. In any case, mixing of the syringe contents by turning the syringe in a top-to-bottom 180° shaking movement 5 times with an air bubble would ensure administration of the drug at a constant concentration.

Conclusions: Without mixing, the concentrations of all drug solutions varied widely when administered via an electronic syringe infusion pump. Mixing syringe contents should be made part of the compulsory curriculum for administering medications at all levels of medical education. (Critical Care Nurse. 2016;36[4]:36-45).

Download full-text PDF

Source
http://dx.doi.org/10.4037/ccn2016756DOI Listing

Publication Analysis

Top Keywords

syringe
9
syringe pump
8
drug
8
drug solutions
8
electronic syringe
8
syringe infusion
8
drug vial
8
vial solvent
8
air bubble
8
turning syringe
8

Similar Publications

Background: Correct identification of the epidural space requires extensive training for technical proficiency. This study explores a novel bimanual haptic simulator designed for the precise insertion of an epidural needle based on loss-of-resistance (LOR) detection, providing realistic dual-hand force feedback.

Methods: The simulator, equipped with two haptic devices connected to a Tuohy needle and an LOR syringe, was designed to simulate the tissues' resistive forces felt by the user during the procedure, offer anatomical variability and record detailed performance metrics for personalized feedback.

View Article and Find Full Text PDF

Tc toxins are pore-forming virulence factors of many pathogenic bacteria. Following pH-induced conformational changes, they perforate the target membrane like a syringe to translocate toxic enzymes into a cell. Although this complex transformation has been structurally well studied, the reaction pathway and the resulting temporal evolution have remained elusive.

View Article and Find Full Text PDF

A simple negative pressure drainage device for skin wounds.

J Am Acad Dermatol

December 2024

Department of Dermatology, the First Affiliated Hospital of Fujian Medical University, 20 Chazhong Road, Fuzhou, Fujian 350000, China;; Key Laboratory of Skin Cancer of Fujian Higher Education Institutions, the First Affiliated Hospital, Fujian Medical University, Fuzhou, Fujian 350000, China;; Fujian Dermatology and Venereology Research Institute, the First Affiliated Hospital, Fujian Medical University, Fuzhou, China. Electronic address:

View Article and Find Full Text PDF

Therapeutic monoclonal antibody (mAb) drug products are increasingly used to treat both chronic and acute diseases. These mAb drug products are often developed for subcutaneous (SC) injection to simplify dosing compared with intravenous (IV) infusion. For SC injection, the mAb liquid drug product is typically filled in a vial for use with a syringe or in a prefilled syringe, which can then be assembled into a safety syringe device or an autoinjector for direct administration.

View Article and Find Full Text PDF

Objective: We aimed to discuss our experience of a higher incidence of catheter-associated bloodstream infections (CLABSIs) during the needle-free connector (NFC) and single-use prefilled flushing syringe (PFS) shortage.

Materials And Methods: Retrospective analyses were carried out to investigate the CLABSI rates at a tertiary training hospital from January 1, 2023, to December 31, 2023, and the study period included a three-month shortage of NFCs and PFSs in April and June 2023.

Results: The CLABSI rate for the three months was 5.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!