Objective: Preparation of drug solutions used with electronic syringe infusion pumps plays a crucial role in the delivery of an accurate drug concentration. Is there a correlation between drug concentrations during syringe pump infusion and preparation protocols?
Method: Norepinephrine, insulin, and sufentanil were prepared in 3 different ways: (1) the drug was taken from the vial, then the solvent was added followed by an air bubble, and mixing was performed by turning the syringe top-to-bottom in a 180° shaking movement 5 consecutive times; (2) the drug was taken from the vial, then the solvent was added and not mixed; and (3) the solvent was taken from a stock solution, then the drug was added and not mixed. Concentrations of drugs were determined at different times during administration by reverse-phase high-performance liquid chromatography with ultraviolet detection. All analyses were performed in triplicate and were based on measurement of peak areas.
Results: With no shaking of the syringe, the concentration of the injected drugs varies widely. In any case, mixing of the syringe contents by turning the syringe in a top-to-bottom 180° shaking movement 5 times with an air bubble would ensure administration of the drug at a constant concentration.
Conclusions: Without mixing, the concentrations of all drug solutions varied widely when administered via an electronic syringe infusion pump. Mixing syringe contents should be made part of the compulsory curriculum for administering medications at all levels of medical education. (Critical Care Nurse. 2016;36[4]:36-45).
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.4037/ccn2016756 | DOI Listing |
Int J Obstet Anesth
December 2024
Department of Biomedical Engineering and the School of Brain Sciences and Cognition, Ben Gurion University of the Negev, Beer Sheva, Israel.
Background: Correct identification of the epidural space requires extensive training for technical proficiency. This study explores a novel bimanual haptic simulator designed for the precise insertion of an epidural needle based on loss-of-resistance (LOR) detection, providing realistic dual-hand force feedback.
Methods: The simulator, equipped with two haptic devices connected to a Tuohy needle and an LOR syringe, was designed to simulate the tissues' resistive forces felt by the user during the procedure, offer anatomical variability and record detailed performance metrics for personalized feedback.
Sci Adv
January 2025
Department of Structural Biochemistry, Max Planck Institute of Molecular Physiology, Otto-Hahn-Str. 11, 44227 Dortmund, Germany.
Tc toxins are pore-forming virulence factors of many pathogenic bacteria. Following pH-induced conformational changes, they perforate the target membrane like a syringe to translocate toxic enzymes into a cell. Although this complex transformation has been structurally well studied, the reaction pathway and the resulting temporal evolution have remained elusive.
View Article and Find Full Text PDFJ Am Acad Dermatol
December 2024
Department of Dermatology, the First Affiliated Hospital of Fujian Medical University, 20 Chazhong Road, Fuzhou, Fujian 350000, China;; Key Laboratory of Skin Cancer of Fujian Higher Education Institutions, the First Affiliated Hospital, Fujian Medical University, Fuzhou, Fujian 350000, China;; Fujian Dermatology and Venereology Research Institute, the First Affiliated Hospital, Fujian Medical University, Fuzhou, China. Electronic address:
Mol Pharm
January 2025
Drug Product Development-Steriles, Medicine Development and Supply, GSK, 1250 S. Collegeville Road, Collegeville, Pennsylvania 19426, United States.
Therapeutic monoclonal antibody (mAb) drug products are increasingly used to treat both chronic and acute diseases. These mAb drug products are often developed for subcutaneous (SC) injection to simplify dosing compared with intravenous (IV) infusion. For SC injection, the mAb liquid drug product is typically filled in a vial for use with a syringe or in a prefilled syringe, which can then be assembled into a safety syringe device or an autoinjector for direct administration.
View Article and Find Full Text PDFInfect Dis Clin Microbiol
December 2024
Department of Pediatric Intensive Care Unit, Dr. Behçet Uz Children's Diseases and Surgery Training and Research Hospital, İzmir, Türkiye.
Objective: We aimed to discuss our experience of a higher incidence of catheter-associated bloodstream infections (CLABSIs) during the needle-free connector (NFC) and single-use prefilled flushing syringe (PFS) shortage.
Materials And Methods: Retrospective analyses were carried out to investigate the CLABSI rates at a tertiary training hospital from January 1, 2023, to December 31, 2023, and the study period included a three-month shortage of NFCs and PFSs in April and June 2023.
Results: The CLABSI rate for the three months was 5.
Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!