Aim: The aim of the present study was to compare the accuracy of radiographic measurements for dental implants planning using cone-beam computed tomography (CBCT) and helical computed tomography (HCT).

Methods: Six pig ribs were wrapped by putty impression material, with radiographic markers placed. Two CBCT and an HCT were taken following the standard protocols. Twenty-five locations were selected, with vertical and horizontal dimensions measured using the default software, as well as on the processed HCT films by a digital caliper. The actual dimensions of the ribs measured by the digital caliper served as the control. Differences between radiographic dimensions and the actual dimensions were tested by two-way analysis of variance.

Results: No differences were found between measurements made by CBCT and HCT images using the default software (P > 0.05). However, both measurements were statistically-significantly lower than the control (P < 0.001), and the mean difference was 0.3 mm. Measurements made on HCT films were statistically-significantly greater than the control (P < 0.001), and the mean difference was 0.5 mm.

Conclusion: The accuracy of CBCT and HCT are similar, and both are reliable tools for implant planning. It is preferable to perform the planning using default software, rather than making direct measurements on films.

Download full-text PDF

Source
http://dx.doi.org/10.1111/jicd.12232DOI Listing

Publication Analysis

Top Keywords

cbct hct
12
default software
12
accuracy radiographic
8
radiographic measurements
8
implant planning
8
planning cone-beam
8
computed tomography
8
hct films
8
digital caliper
8
actual dimensions
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!