Temporal artifact minimization in sonoelastography through optimal selection of imaging parameters.

J Acoust Soc Am

Laboratorio de Imágenes Médicas, Departamento de Ingeniería, PontificiaUniversidad Católica del Perú, Lima 32, Peru.

Published: July 2016

Sonoelastography is an ultrasonic technique that uses Kasai's autocorrelation algorithms to generate qualitative images of tissue elasticity using external mechanical vibrations. In the absence of synchronization between the mechanical vibration device and the ultrasound system, the random initial phase and finite ensemble length of the data packets result in temporal artifacts in the sonoelastography frames and, consequently, in degraded image quality. In this work, the analytic derivation of an optimal selection of acquisition parameters (i.e., pulse repetition frequency, vibration frequency, and ensemble length) is developed in order to minimize these artifacts, thereby eliminating the need for complex device synchronization. The proposed rule was verified through experiments with heterogeneous phantoms, where the use of optimally selected parameters increased the average contrast-to-noise ratio (CNR) by more than 200% and reduced the CNR standard deviation by 400% when compared to the use of arbitrarily selected imaging parameters. Therefore, the results suggest that the rule for specific selection of acquisition parameters becomes an important tool for producing high quality sonoelastography images.

Download full-text PDF

Source
http://dx.doi.org/10.1121/1.4958997DOI Listing

Publication Analysis

Top Keywords

optimal selection
8
imaging parameters
8
ensemble length
8
selection acquisition
8
acquisition parameters
8
parameters
5
temporal artifact
4
artifact minimization
4
sonoelastography
4
minimization sonoelastography
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!