False intensive care unit (ICU) alarms induce stress in both patients and clinical staff and decrease the quality of care, thus significantly increasing both the hospital recovery time and rehospitalization rates. In the PhysioNet/CinC Challenge 2015 for reducing false arrhythmia alarms in ICU bedside monitor data, this paper validates the application of a real-time arrhythmia detection library (ADLib, Schiller AG) for the robust detection of five types of life-threatening arrhythmia alarms. The strength of the application is to give immediate feedback on the arrhythmia event within a scan interval of 3 s-7.5 s, and to increase the noise immunity of electrocardiogram (ECG) arrhythmia analysis by fusing its decision with supplementary ECG quality interpretation and real-time pulse wave monitoring (quality and hemodynamics) using arterial blood pressure or photoplethysmographic signals. We achieved the third-ranked real-time score (79.41) in the challenge (Event 1), however, the rank was not officially recognized due to the 'closed-source' entry. This study shows the optimization of the alarm decision module, using tunable parameters such as the scan interval, lead quality threshold, and pulse wave features, with a follow-up improvement of the real-time score (80.07). The performance (true positive rate, true negative rate) is reported in the blinded challenge test set for different arrhythmias: asystole (83%, 96%), extreme bradycardia (100%, 90%), extreme tachycardia (98%, 80%), ventricular tachycardia (84%, 82%), and ventricular fibrillation (78%, 84%). Another part of this study considers the validation of ADLib with four reference ECG databases (AHA, EDB, SVDB, MIT-BIH) according to the international recommendations for performance reports in ECG monitors (ANSI/AAMI EC57). The sensitivity (Se) and positive predictivity (+P) are: QRS detector QRS (Se, +P)  >  99.7%, ventricular ectopic beat (VEB) classifier VEB (Se, +P)  =  95%, and ventricular fibrillation detector VFIB (P  +  = 94.8%)  >  VFIB (Se  =  86.4%), adjusted to the clinical setting requirements, giving preference to low false positive alarms.

Download full-text PDF

Source
http://dx.doi.org/10.1088/0967-3334/37/8/1273DOI Listing

Publication Analysis

Top Keywords

pulse wave
12
real-time arrhythmia
8
arrhythmia detection
8
supplementary ecg
8
ecg quality
8
wave monitoring
8
arrhythmia alarms
8
scan interval
8
real-time score
8
ventricular fibrillation
8

Similar Publications

There is scant information available about the blood flow of the pulmonary artery in avian cardiology. In human medicine, the shape of the Doppler sonographic blood flow profile of the pulmonary artery can be used to access the pressure conditions of the right heart. With this background, this study focused on the examination of the acceleration and deceleration phase of the pulsed-wave Doppler flow profile of the pulmonary artery of healthy racing pigeons.

View Article and Find Full Text PDF

Foods rich in polyphenols have beneficial effects on health. This study aimed to examine the impact of dark chocolate on endurance runners' arterial function. Forty-six male amateur runners, aged 25-55, participated.

View Article and Find Full Text PDF

Trimethylamine -oxide (TMAO), a gut microbiome-derived metabolite, participates in the atherogenesis and vascular stiffening that is closely linked with cardiovascular (CV) complications and related deaths in individuals with kidney failure undergoing peritoneal dialysis (PD) therapy. In these patients, arterial stiffness (AS) is also an indicator of adverse CV outcomes. This study assessed the correlation between serum TMAO concentration quantified with high-performance liquid chromatography and mass spectrometry and central AS measured by carotid-femoral pulse wave velocity (cfPWV) in patients with chronic PD.

View Article and Find Full Text PDF

: Controlling hypertension may reduce the risk of cognitive impairment. A marker for the identification of hypertensive patients who are more likely to suffer cognitive impairment would be of clinical benefit. In our research, 105 patients with newly diagnosed primary hypertension were assessed at the Department of Neurology, the University of Debrecen.

View Article and Find Full Text PDF

Background: This study explored the potential of electrogastrography (EGG) and heart rate variability (HRV) as psychophysiological markers in experimental pain research related to the gut-brain axis. We investigated responses to the experience of pain from the visceral (rectal distension) and somatic (cutaneous heat) pain modalities, with a focus on elucidating sex differences in EGG and HRV responses.

Methods: In a sample of healthy volunteers (29 males, 43 females), EGG and ECG data were collected during a baseline and a pain phase.

View Article and Find Full Text PDF