Practical improvements in the design of high permittivity pads for dielectric shimming in neuroimaging at 7T.

J Magn Reson

C.J. Gorter Center for High Field MRI, Department of Radiology, Leiden University Medical Center, Leiden, Netherlands.

Published: September 2016

Improvements are proposed for practical design and use of high permittivity materials in high field neuroimaging in three different areas: (i) a simple formula to predict the permittivity of tri-component aqueous-based perovskite suspensions with relative permittivities between 110 and 300, (ii) characterization of addition of a hydroxyethyl-cellulose gelling agent to improve the long-term stability and material properties of "dielectric pads", and (iii) investigation of the integration of, for example, headphones into the dielectric pads to increase patient comfort within tightly-fitting receive coil arrays.