Purpose: For the stabilization of the thoracolumbar spine area, various stabilization techniques have been developed in recent decades. The aim of these techniques is to immobilize the treated segment to repositioning or correct the spine and guaranty long-term stability to achieve a reliable fusion. The aim of this study was to simulate in an in vitro experiment the postoperative long-term situation in elderly osteoporotic patients to compare two different stabilization principles; a pedicle screw system and a lamina hook system.

Methods: Two comparable groups with respect to age and bone mineral density with each n = 6 fresh-frozen human, bi-segmental thoracolumbar spine specimens (T11-L1) were used. Antero-posterior and lateral radiographs were taken before the test, to assess the spinal status. Then the intact specimens were biomechanically characterized with pure moments in the three anatomical planes in different states in terms of range of motion and neutral zone. After implantation of either, a pedicle screw system or a lamina hook system, the primary stability was determined under the same conditions. Subsequently the specimens were cyclically loaded under complex loading, using a custom-made set-up in a dynamic materials testing machine with increasing moments from 3 to 66 Nm until 100,000 cycles or until one of the three defined "failure" criteria was reached. (1) A failure of a bony structure. (2) Exceeding of the threefold ROM of the primary stability after implantation in flexion plus extension. (3) Reaching of the ROM based on the intact state before implantation both in flexion plus extension.

Results: The results showed that the ROM was strongly reduced after instrumentation similar for both implant systems in all motion planes. The highest stabilization was found in flexion/extension. During cyclic loading with increasing moments, the ROM increased continuously for both systems. The number of load cycles until one of the failure criteria was reached varied only slightly between the two groups. In the pedicle screw group 30,000 (median) loading cycles (range 5000-80,000) with a corresponding moment of 24 Nm (range 9-54) could be reached. In the lamina hook group 32,500 load cycles (range 20,000-45,000) could be achieved with a corresponding moment of 25.5 Nm (range 18-33). There was a slight trend that the pedicle screw system is influenced more by bone mineral density.

Conclusion: Both implant systems provide similar primary stability and similar long-term stability. In the pedicle screw group, there was a stronger correlation between bone mineral density and the reached number of load cycles.

Download full-text PDF

Source
http://dx.doi.org/10.1007/s00586-016-4679-xDOI Listing

Publication Analysis

Top Keywords

pedicle screw
24
screw system
16
lamina hook
16
system lamina
12
long-term stability
12
bone mineral
12
primary stability
12
load cycles
12
hook system
8
provide primary
8

Similar Publications

Preserving Cervical Mobility: A Novel Robot-Assisted Approach for Atlas Fracture Fixation.

Am J Case Rep

January 2025

Department of Orthopedic Surgery, Sichuan Provincial People's Hospital, School of Medicine, University of Electronic Science and Technology of China, Chengdu, Sichuan, China.

BACKGROUND The management of unstable atlas fractures remains a subject of ongoing debate and controversy. The conservative surgical treatment commonly involves fusion, resulting in severe loss of cervical spine mobility, and a large incisions and extensive tissue dissection are required. We aim to introduce a novel concept and surgical approach for treating atlas fracture, one that involves minimizing trauma while maintaining mobility of the upper cervical spine without resorting to fusion.

View Article and Find Full Text PDF

Objective: To observe the clinical outcomes of anterior approach for the revision surgery following unsuccessful bone cement augmentation in osteoporotic vertebral compression fractures.

Methods: A total of 10 patients who experienced unsuccessful bone cement augmentation underwent anterior revision surgery between January 2020 and December 2021. There were 2 males and 8 females.

View Article and Find Full Text PDF

Delayed diagnosis of spinal osteoblastoma presenting with radicular pain and scoliosis: A case report.

Int J Surg Case Rep

January 2025

University of Tunis El Manar, Faculty of Medicine of Tunis, 1007, Tunisia; Department of Orthopedic Surgery, Hospital Mongi Slim La Marsa, Tunisia.

Introduction And Importance: Osteoblastoma is a rare benign bone tumor, accounting for 1 % of primary bone tumors, often affecting the spine and sacrum. Accurate diagnosis is essential for appropriate treatment and prognosis.

Case Presentation: A 19-year-old male presented with two years of persistent nocturnal radicular and low back pain unresponsive to anti-inflammatory medications.

View Article and Find Full Text PDF

Objectives: The advent of O-arm navigation optimized the oblique lumbar interbody fusion (OLIF) procedure, allowing the operator to simultaneously perform OLIF and percutaneous posterior pedicle screw implantation without patient position change, thus improving the fluency and accuracy of the OLIF procedure (called as OLIF360). Nevertheless, a consensus regarding its suitability for patients with severe spinal stenosis remains elusive. This study aims to investigate the clinical efficacy of OLIF360 and its imaging changes in severe lumbar spinal stenosis cases.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!