Purpose: To evaluate the accuracy of R2* models (1/T * = R2*) for chemical shift-encoded magnetic resonance imaging (CSE-MRI)-based proton density fat-fraction (PDFF) quantification in patients with fatty liver and iron overload, using MR spectroscopy (MRS) as the reference standard.
Materials And Methods: Two Monte Carlo simulations were implemented to compare the root-mean-squared-error (RMSE) performance of single-R2* and dual-R2* correction in a theoretical liver environment with high iron. Fatty liver was defined as hepatic PDFF >5.6% based on MRS; only subjects with fatty liver were considered for analyses involving fat. From a group of 40 patients with known/suspected iron overload, nine patients were identified at 1.5T, and 13 at 3.0T with fatty liver. MRS linewidth measurements were used to estimate R2* values for water and fat peaks. PDFF was measured from CSE-MRI data using single-R2* and dual-R2* correction with magnitude and complex fitting.
Results: Spectroscopy-based R2* analysis demonstrated that the R2* of water and fat remain close in value, both increasing as iron overload increases: linear regression between R2* and R2* resulted in slope = 0.95 [0.79-1.12] (95% limits of agreement) at 1.5T and slope = 0.76 [0.49-1.03] at 3.0T. MRI-PDFF using dual-R2* correction had severe artifacts. MRI-PDFF using single-R2* correction had good agreement with MRS-PDFF: Bland-Altman analysis resulted in -0.7% (bias) ± 2.9% (95% limits of agreement) for magnitude-fit and -1.3% ± 4.3% for complex-fit at 1.5T, and -1.5% ± 8.4% for magnitude-fit and -2.2% ± 9.6% for complex-fit at 3.0T.
Conclusion: Single-R2* modeling enables accurate PDFF quantification, even in patients with iron overload.
Level Of Evidence: 1 J. Magn. Reson. Imaging 2017;45:428-439.
Download full-text PDF |
Source |
---|---|
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC5420327 | PMC |
http://dx.doi.org/10.1002/jmri.25382 | DOI Listing |
Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!