Background: Blood product safety was significantly improved by the introduction of NAT testing in the late 1990s, resulting in a strong decrease of transfusion-transmitted infections (TTIs). Due to the occurrence of HIV-1 NAT test failures as a consequence of mismatch mutations in the amplicon regions of mono-target NAT assays, the Paul Ehrlich Institute mandated the implementation of multi-target NAT assays for HIV-1 in 2014. Commercial suppliers mostly developed dual-target NAT assays, with only one implementing a triple-target NAT assay.
Methods: The HIV-1 triple-target NAT assay v3 (GFE Blut) was tested on mutated specimens and synthetic DNA bearing mutations that resulted in sample underquantification or false-negative test results. In addition, data from 2 years routine testing at three German Red Cross Blood centres were analysed.
Results: The HIV-1 triple-target PCR could compensate for all mutations tested and could compensate the loss of one amplicon without a significant loss of sensitivity. Data from 2 years routine testing showed a solid performance.
Conclusion: The HIV-1 triple-target v3 assay (GFE Blut) can compensate mutations in target sequences better than a dual-target assay and is applicable to high-throughput screening, thus increasing blood product safety.
Download full-text PDF |
Source |
---|---|
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC4924456 | PMC |
http://dx.doi.org/10.1159/000446290 | DOI Listing |
Transfus Med Hemother
May 2016
Gesellschaft zur Forschung, Entwicklung und Distribution von Diagnostika im Blutspendewesen mbH (GFE Blut), Frankfurt/M., Germany.
Background: Blood product safety was significantly improved by the introduction of NAT testing in the late 1990s, resulting in a strong decrease of transfusion-transmitted infections (TTIs). Due to the occurrence of HIV-1 NAT test failures as a consequence of mismatch mutations in the amplicon regions of mono-target NAT assays, the Paul Ehrlich Institute mandated the implementation of multi-target NAT assays for HIV-1 in 2014. Commercial suppliers mostly developed dual-target NAT assays, with only one implementing a triple-target NAT assay.
View Article and Find Full Text PDFEnter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!