Background: Blood product safety was significantly improved by the introduction of NAT testing in the late 1990s, resulting in a strong decrease of transfusion-transmitted infections (TTIs). Due to the occurrence of HIV-1 NAT test failures as a consequence of mismatch mutations in the amplicon regions of mono-target NAT assays, the Paul Ehrlich Institute mandated the implementation of multi-target NAT assays for HIV-1 in 2014. Commercial suppliers mostly developed dual-target NAT assays, with only one implementing a triple-target NAT assay.

Methods: The HIV-1 triple-target NAT assay v3 (GFE Blut) was tested on mutated specimens and synthetic DNA bearing mutations that resulted in sample underquantification or false-negative test results. In addition, data from 2 years routine testing at three German Red Cross Blood centres were analysed.

Results: The HIV-1 triple-target PCR could compensate for all mutations tested and could compensate the loss of one amplicon without a significant loss of sensitivity. Data from 2 years routine testing showed a solid performance.

Conclusion: The HIV-1 triple-target v3 assay (GFE Blut) can compensate mutations in target sequences better than a dual-target assay and is applicable to high-throughput screening, thus increasing blood product safety.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC4924456PMC
http://dx.doi.org/10.1159/000446290DOI Listing

Publication Analysis

Top Keywords

hiv-1 triple-target
16
triple-target nat
12
nat assays
12
nat
8
nat assay
8
three german
8
german red
8
red cross
8
cross blood
8
blood centres
8

Similar Publications

Implementation of an HIV-1 Triple-Target NAT Assay in the Routine Screening at Three German Red Cross Blood Centres.

Transfus Med Hemother

May 2016

Gesellschaft zur Forschung, Entwicklung und Distribution von Diagnostika im Blutspendewesen mbH (GFE Blut), Frankfurt/M., Germany.

Background: Blood product safety was significantly improved by the introduction of NAT testing in the late 1990s, resulting in a strong decrease of transfusion-transmitted infections (TTIs). Due to the occurrence of HIV-1 NAT test failures as a consequence of mismatch mutations in the amplicon regions of mono-target NAT assays, the Paul Ehrlich Institute mandated the implementation of multi-target NAT assays for HIV-1 in 2014. Commercial suppliers mostly developed dual-target NAT assays, with only one implementing a triple-target NAT assay.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!