Background: The mini C-arm is popular with hand surgeons, and they are particularly at risk for radiation exposure, as they cannot easily distance themselves from the radiation beam. We hypothesized that the nondominant hand receives more radiation exposure than the dominant hand as it is generally closer to the radiation source. This study seeks to determine whether a hand surgeon receives a different amount of radiation exposure to their hands based on hand dominance and then accounting for this, provide a more accurate assessment of hand radiation exposure from mini C-arm fluoroscopy.

Methods: Two fellowship-trained hand surgeons wore ring dosimeters on both hands during surgeries with mini C-arm fluoroscopy involving bony work of the forearm and hand. Radiation exposure to the hands was measured from ring dosimeters.

Results: Six-ring dosimeter pairs were worn during 64 cases, averaging 10.7 cases per ring. No ring met the minimal dose threshold of 30 mrem to record a numerical value. Each ring experienced an average of 413 seconds of fluoroscopy time and 66.3509 cGy*cm(2) of radiation output from the mini C-arm.

Conclusions: The results do not allow comparison of radiation exposure related to hand dominance. Assuming worst-case scenario: each ring measured 29 mrem (just below the threshold), the surgeon's hands experienced 2.7 mrem per case. This would allow a hand surgeon to perform 18 391 cases per year before exceeding the allowable annual hand exposure limit of 50 000 mrem set by the National Council of Radiation Protection and Measurements and International Commission on Radiological Protection.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC4920527PMC
http://dx.doi.org/10.1177/1558944715627224DOI Listing

Publication Analysis

Top Keywords

radiation exposure
28
mini c-arm
16
hand
13
hand dominance
12
radiation
11
exposure hand
8
c-arm fluoroscopy
8
hand surgeons
8
hand surgeon
8
exposure hands
8

Similar Publications

Cancer remains a leading cause of mortality worldwide, requiring physicians to understand multidisciplinary treatments. This study assessed the impact of a clinical rotation in a cancer center on medical students' knowledge of cancer treatments from a multidisciplinary perspective. A traditional single-department rotation was compared to a multidisciplinary rotation to determine whether broader exposure enhances knowledge and prepares students for multidisciplinary care.

View Article and Find Full Text PDF

Effect of terahertz radiation on cells and cellular structures.

Front Optoelectron

January 2025

Institute of Physics, Saratov State University, Saratov, 410012, Russia.

The paper presents the results of modern research on the effects of electromagnetic terahertz radiation in the frequency range 0.5-100 THz at different levels of power density and exposure time on the viability of normal and cancer cells. As an accompanying tool for monitoring the effect of radiation on biological cells and tissues, spectroscopic research methods in the terahertz frequency range are described, and attention is focused on the possibility of using the spectra of interstitial water as a marker of pathological processes.

View Article and Find Full Text PDF

Objective: To explore the efficacy and safety of the direct inferior endplate approach in percutaneous endoscopic interlaminar discectomy (PEID) for the treatment of L5-S1 disc herniation.

Methods: This was a retrospective analysis of 116 patients with L5-S1 disc herniation treated with PEID; 74 patients underwent surgery via the direct inferior endplate approach (group A), and 42 patients underwent surgery via the indirect approach (group B). The number of intraoperative fluoroscopy exposures, establishment channel time, operation time, postoperative visual analogue scale (VAS) score, and Oswestry Disability Index (ODI) were compared between the 2 groups.

View Article and Find Full Text PDF

Objectives: In advanced stages of osteoradionecrosis, medication-related osteonecrosis of the jaw, and osteomyelitis, a resection of sections of the mandible may be unavoidable. The determination of adequate bony resection margins is a fundamental problem because bony resection margins cannot be secured intraoperatively. Single-photon emission computed tomography (SPECT-CT) is more accurate than conventional imaging techniques in detecting inflammatory jaw pathologies.

View Article and Find Full Text PDF

Keratin/chitosan film promotes wound healing in rats with combined radiation-wound injury.

J Mater Sci Mater Med

January 2025

Department of Nuclear Medicine, Chongqing University Cancer Hospital, No. 181 HanYu St, Shapingba District, Chongqing, 400030, PR China.

Human hair keratin, a natural protein derived from human hair, has emerged prominently in the field of wound repair, showcasing its unique regenerative capabilities and extensive application potential. However, it is a challenge for the keratin to efficiently therapy the impaired wound healing, such as combined radiation-wound injury. Here, we report a keratin/chitosan (KRT/CS) film for skin repair of chronic wounds in in rats with combined radiation-wound injury.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!