Periprosthetic femoral fractures are a common complication associated with cementless press-fit total hip arthroplasty. The use of prophylactic cerclage wire fixation has been advocated to reduce this complication. The objective of this study was to evaluate whether a double loop cerclage wire, used as adjunctive fixation, increased the peak torsional load to failure in femora implanted with press-fit cementless stems. Peak torsional load to failure was compared between femora without adjunctive fixation and femora receiving a 1 mm double loop cerclage wire placed proximally to the lesser trochanter. Femora treated with adjunctive cerclage wire fixation failed at 20% greater peak torque (P = 0.0001). In conclusion, a double loop cerclage wire may aid in the prevention of periprosthetic fractures associated with press-fit cementless femoral stems.