Background: An implantable cardioverter-defibrillator (ICD) is routinely used to prevent sudden cardiac death. Since the introduction of that device into clinical practice, a defibrillation test (the so-called pre-discharge test [PDT]) has been an inseparable part of the ICD implantation procedure. Recently, the usefulness of PDT has been called into question.

Methods: The aim of this research was to analyze ICD tests performed within two time periods: in years 1995-2001 (period I) and 2007-2010 (period II), in order to compare the results of tests and solutions to all the problems with ICD systems revealed by means of PDT.

Results: During period I, 193 tests were performed, among which the ICD system malfunction was observed in 16 cases. Those included: sensing issues, specifically R-wave undersensing during ventricular fibrillation (VF) (7 patients) and T-wave oversensing (4 patients), as well as high defibrillation threshold (DFT) (2 patients) and ICD-pacemaker interaction (3 patients). During period II, among 561 tests, system malfunction was observed in 15 cases. In 1 patient it was VF undersensing, and in the remaining 14 it was high DFT. All the above problems were solved by means of appropriate ICD reprogramming, repositioning of the endocardial defibrillation lead or implantation of an additional subcutaneous defibrillation lead.

Conclusions: Contemporary ICD technical solutions, compared to older systems, in most cases allow to avoid sensing problems. The key rationale behind ICD testing is the ability to confirm the efficacy of high-voltage therapy. Despite the increasing maximal defibrillation out-put of devices, and all possible adjustments to the characteristics of the impulse, there is still a group of patients that require additional procedures to ensure the appropriate defibrillation efficacy.

Download full-text PDF

Source
http://dx.doi.org/10.5603/CJ.a2016.0045DOI Listing

Publication Analysis

Top Keywords

icd
8
tests performed
8
system malfunction
8
malfunction observed
8
observed cases
8
defibrillation
6
patients
5
pre-discharge defibrillation
4
defibrillation testing
4
testing justified?
4

Similar Publications

Camellia seed oil (CSO), a potential prebiotic agent, can significantly increase the relative abundance of () in mice gut microbiota following oral administration, this study aims to investigate the enhancing effect in vitro. The results showed that after 24-h co-cultivation with 0.5% (v/v) CSO, the growth of increased from 11.

View Article and Find Full Text PDF

Cardiac implantable electronic devices (CIEDs), including pacemakers, implantable cardiac defibrillators (ICD), and cardiac resynchronization therapy (CRT) devices, regulate heart rate and rhythm in patients with cardiac conditions. With an aging population, CIED-related complications, especially pacemaker pocket infections, are rising. Risk factors include frailty, older age, and superficial device fixation, while risk mitigation involves larger pocket sizes, submuscular fixation, and absorbable antibacterial envelopes.

View Article and Find Full Text PDF

Objective: To examine the association between mood disorders in pregnancy and postpartum and peripartum cardiomyopathy (PPCM).

Methods: Retrospective cohort study utilizing the National Inpatient Sample from the Healthcare Cost and Utilization Project, Agency for Healthcare Research and Quality of pregnant and postpartum patients from 2017-2019. Patients were separated into two groups based on ICD-10 coding for presence or absence of mood disorder (depression, bipolar depression, anxiety, or other mood diagnosis).

View Article and Find Full Text PDF

Purpose: Recent redefinitions of pain emphasize the importance of the previously overlooked recurrent low back pain (LBP). Understanding the direct medical cost for recurrent LBP cases based on the cost per visit is crucial economically. We aimed to compare the cost per visit for LBP and recurrent LBP, including the impact of gender and type of medical service, estimating the approximate annual cost of recurrent LBP.

View Article and Find Full Text PDF

Mitochondria-Targeting Type-I Photodynamic Therapy Based on Phenothiazine for Realizing Enhanced Immunogenic Cancer Cell Death via Mitochondrial Oxidative Stress.

Int J Nanomedicine

January 2025

Guangzhou Institute of Cancer Research, the Affiliated Cancer Hospital, Guangzhou Medical University, Guangzhou, Guangdong, 510095, People's Republic of China.

Purpose: Photo-immunotherapy faces challenges from poor immunogenicity and low response rate due to hypoxic microenvironment. This study presents Rh-PTZ, a small organic molecule with a D-π-A structure, that simultaneously amplifies mitochondria-targeted type-I PDT-dependent immune stimulation for the treatment of hypoxic cancer.

Methods: The hydrophobic Rh-PTZ was encapsulated into F127 to prepare Rh-PTZ nanoparticles (Rh-PTZ NPs).

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!