Purpose: To characterize and optimize F MRI for different perfluorocarbons (PFCs) at 3T and quantify the loss of acquisition efficiency as a function of different temperature and cellular conditions.

Methods: The T and T relaxation times of the commonly used PFCs perfluoropolyether (PFPE), perfluoro-15-crown-5-ether (PFCE), and perfluorooctyl bromide (PFOB) were measured in phantoms and in several different conditions (cell types, presence of fixation agent, and temperatures). These relaxation times were used to optimize pulse sequences through numerical simulations. The acquisition efficiency in each cellular condition was then determined as the ratio of the signal after optimization with the reference relaxation times and after optimization with its proper relaxation times. Finally, PFC detection limits were determined.

Results: The loss of acquisition efficiency due to parameter settings optimized for the wrong temperature and cellular condition was limited to 13%. The detection limits of all PFCs were lower at 24 °C than at 37 °C and varied from 11.8 ± 3.0 mM for PFCE at 24 °C to 379.9 ± 51.8 mM for PFOB at 37 °C.

Conclusion: Optimizing F pulse sequences with a known phantom only leads to moderate loss in acquisition efficiency in cellular conditions that might be encountered in in vivo and in vitro experiments. Magn Reson Med 77:2263-2271, 2017. © 2016 International Society for Magnetic Resonance in Medicine.

Download full-text PDF

Source
http://dx.doi.org/10.1002/mrm.26317DOI Listing

Publication Analysis

Top Keywords

relaxation times
20
acquisition efficiency
16
loss acquisition
12
temperature cellular
8
pulse sequences
8
efficiency cellular
8
cellular condition
8
detection limits
8
relaxation
5
times
5

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!