Background: The small diameter of radial arteries remains a major limitation of the transradial approach for percutaneous coronary intervention (PCI). Sheathless guiding catheters (GCs) might offer an advantage over standard GCs.
Methods: Between 2011 and 2013, we randomized 233 transradial PCIs performed in men with ostial or bifurcation lesions and in all women between standard GC (Medtronic Launcher; Minneapolis, MN) and the SheathLess Eaucath GC (Asahi Intecc, Aichi, Japan).
Results: Successful PCI using the transradial approach was not different between the groups (P = 0.74), however the rate of successful transradial PCI with the designated GC (ie, without crossover to the opposing GC) was superior in the SheathLess group compared with the standard GC group (96.5% vs 89.9%; P = 0.047). Safety end point (ie, absence of PCI complication, radial artery occlusion, perforation, pseudoaneurysm, and Early Discharge after Transradial Stenting of Coronary Arteries [EASY] hematoma grade ≥ 2) did not differ between the groups (60.5% in both groups). Mean PCI duration (45.1 minutes vs 45.9 minutes), fluoroscopy (20.1 minutes and 19.9 minutes), and cannulation times (3.6 minutes vs 3.7 minutes), contrast media volume (196 mL vs 187 mL) and conversion to transfemoral approach (1.8% vs 0.8%) were not different between the groups. Patients' subjective assessment revealed less arm pain during navigation of the SheathLess GC (1.9 ± 1.9 vs 4.8 ± 3.6; P < 0.001). Operators graded arm crossability as easier with the SheathLess GC (8.7 ± 1.5 vs 5.1 ± 3.5; P < 0.001).
Conclusions: In selected coronary lesions requiring large-bore catheters in men and in all lesions in women, the SheathLess GC was superior to the standard GC for successful transradial PCI with the designated GC. The SheathLess GC was also associated with easier arm navigation and less patient discomfort.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1016/j.cjca.2016.03.016 | DOI Listing |
Micromachines (Basel)
January 2025
School of Mechanical and Electrical Engineering, China University of Mining and Technology, Xuzhou 221116, China.
Inertial microfluidics, as an efficient method for the manipulation of micro-/nanoparticles, has garnered significant attention due to its advantages of high throughput, structural simplicity, no need for external fields, and sheathless operation. Common structures include straight channels, contraction-expansion array (CEA) channels, spiral channels, and serpentine channels. In this study, we developed a CEA channel embedded with hook-shaped microstructures to modify the characteristics of vortices.
View Article and Find Full Text PDFElectrophoresis
January 2025
Department of Toxicology, Advanced Medical and Dental Institute, Universiti Sains Malaysia, Bertam, Kepala Batas, Pulau Pinang, Malaysia.
Advancements in food technology have increased the need for thorough analysis to ensure food safety, quality, and compliance with regulatory requirements. Capillary electrophoresis-mass spectrometry (CE-MS) has emerged as a powerful tool in food analysis due to its high separation efficiency, low sample consumption, and ability to handle complex matrices. However, challenges such as the use of volatile running buffers and maintaining the stability of the electrical circuit connecting the CE and MS systems have been addressed through advancements in interface designs, such as sheathless systems and optimized sheath-liquid compositions.
View Article and Find Full Text PDFCatheter Cardiovasc Interv
December 2024
Department of Cardiology, Hôpital de Jolimont, La Louvrière, Belgium.
Background: Distal trans-radial access (dTRA) for percutaneous coronary interventions (PCI) is increasingly gaining attention due to its potential to mitigate radial artery occlusion (RAO). However, a comprehensive understanding of the mechanical impact of the devices on the radial artery (RA) wall remains limited. Using a complete intravascular ultrasound (IVUS) evaluation of the RA, including also the vascular access site, we aimed to evaluate all the consequences related to the catheterization on the RA wall, starting from the vascular access, comparing conventional sheath and sheathless approaches.
View Article and Find Full Text PDFTalanta
January 2025
Institute of Mass Spectrometry, Zhejiang Engineering Research Center of Advanced Mass Spectrometry and Clinical Application, School of Material Science and Chemical Engineering, Ningbo University, Ningbo, 315211, PR China; Zhenhai Institute of Mass Spectrometry, Ningbo, 315211, PR China; School of Material Science and Chemical Engineering, Ningbo University, Ningbo, 315211, PR China. Electronic address:
A robust interface for coupling capillary electrophoresis (CE) to mass spectrometry (MS) was critical to maintain high separation efficiency of CE while achieving high sensitivity of MS. Current interfaces often suffer from problems such as reproducibility and ruggedness. For this purpose, a new polymetallic-coated sheathless interface was developed for the coupling of CE with MS.
View Article and Find Full Text PDFNeurosurg Rev
October 2024
Department of Endovascular Neurosurgery, Saitama Medical University International Medical Center, 1397-1 Yamane, Hidaka City, Saitama Prefecture, 350-1298, Japan.
Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!