Stereo-electro-encephalography-Guided Radiofrequency Thermocoagulation: From In Vitro and In Vivo Data to Technical Guidelines.

World Neurosurg

Department of Neurosurgery, Hospital for Neurology and Neurosurgery Pierre Wertheimer, Hospices Civils de Lyon, Lyon, France; Université de Lyon, Université Claude Bernard, Lyon, France; Neuroscience Research Center of Lyon, Lyon, France.

Published: October 2016

Background: Deep brain electrodes have been used for the past 10 years to produce bipolar stereo-electro-encephalography-guided radiofrequency thermocoagulation (SEEG RF-TC). However, this technique is based on empiric knowledge. The aim of this study is 3-fold: 1) provide in vivo animal data concerning the effect of bipolar RF-TC on brain and its safety; 2) assess the parameters of this procedure (current delivery and dipole selection) that produce the most efficient lesion; and 3) provide technical guidelines.

Methods: First we achieved in vivo RF-TC on rabbit brains with several conditions (power delivered and lesioning duration) and analyzed their influence on the lesion produced. Only a difference in terms of volume was found, and type of histologic lesions was similar whatever the settings were. We then performed multiple RF-TC in vitro on egg albumen, first with several parameters of radiofrequency and then with different dipole spatial selections. The end point was the size of the radiofrequency thermolesion produced.

Results: Using unfixed parameters of radiofrequency current delivery and increasing it until the power delivered by the generator collapsed produced significantly larger lesions (P = 0.008) than other conditions. Concerning the dipole selection, the use of contiguous contacts on electrodes led to lesions with a higher volume (P = 7.7 × 10) than those produced with noncontiguous ones.

Conclusion: Besides the target selection in SEEG RF-TC, which is summarized on the basis of a literature review, we report the optimal parameters: Radiofrequency current must be increased until the power delivered collapses, and dipoles should be constituted by contiguous electrode contacts.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.wneu.2016.06.095DOI Listing

Publication Analysis

Top Keywords

power delivered
12
parameters radiofrequency
12
stereo-electro-encephalography-guided radiofrequency
8
radiofrequency thermocoagulation
8
seeg rf-tc
8
current delivery
8
dipole selection
8
radiofrequency current
8
radiofrequency
5
rf-tc
5

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!