[Evaluation of the measurement of capillary glucose concentration versus plasma glucose in the newborn].

Arch Pediatr

Service de réanimation et de médecine néonatale, centre de maternité et de néonatalogie, EPS Fattouma Bourguiba, 5000 Monastir, Tunisie; CHU Fattouma Bourguiba, Monastir, faculté de médecine de Monastir, EPS Fattouma Bourguiba, 5000 Monastir, Tunisie.

Published: September 2016

Background: The reliability of blood glucose monitoring in neonatology is not always confirmed. The aim of this study was to evaluate the reliability of blood glucose measurements made with three different devices in newborns.

Patients And Methods: The study was prospective, conducted in a medical and neonatal intensive care department over a period of 4 months. Capillary glucose level was measured with three different glucometers and compared with venous glucose level determined using the hexokinase method. An ANOVA and Scheffe test were used for the correlation analysis.

Results: Three hundred and nine infants were included, with a mean age of 55h and a mean term of 39 weeks of gestation. Mean blood glucose in the laboratory was 0.62±0.15g/L, 0.71±0.17g/L for Accu-Chek(®) Active, 0.80±0.17g/L for Accu-Chek(®) Performa, and 0.83±0.12g/L for Bionime. An ANOVA showed statistically significant differences between the measurements made by glucometers compared to the reference blood glucose levels, and the Scheffé method showed that glucometers overestimated the real plasma glucose levels.

Conclusion: None of the devices used in this study was satisfactory. However, an estimation of blood glucose taking into consideration this numerical overestimation would allow early detection of hypoglycemia.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.arcped.2016.04.025DOI Listing

Publication Analysis

Top Keywords

blood glucose
20
glucose
10
capillary glucose
8
plasma glucose
8
reliability blood
8
glucose level
8
glucometers compared
8
blood
5
[evaluation measurement
4
measurement capillary
4

Similar Publications

To evaluate the effectiveness and feasibility of the copper bianstone scraping combined with Chinese modified termination hypertension dietary therapy program by comparing and analyzing the improvement of blood pressure, blood lipids and blood glucose in hypertensive patients who received copper bianstone scraping combined with Chinese modified termination hypertension dietary therapy intervention. We selected 160 cases of hypertensive patients from July 2022 to March 2024 for the study. They were divided into 80 cases in the comparison group and 80 cases in the observation group according to whether or not they underwent copper bianstone scraping combined with Chinese modified dietary therapy for termination of hypertension.

View Article and Find Full Text PDF

Background: This study evaluates the efficacy and safety of sitagliptin versus gliclazide, combined with metformin, in treatment-naive patients with type 2 diabetes mellitus (T2DM) and glucotoxicity.

Methods: In this single-center, randomized, controlled noninferiority trial, 129 treatment-naive patients with T2DM with glucotoxicity (fasting plasma glucose [FPG] ≥ 200 mg/dL and glycated hemoglobin ≥ 9.0%) were randomized to receive sitagliptin plus metformin (n = 66) or gliclazide plus metformin (n = 63) for 12 weeks.

View Article and Find Full Text PDF

Gitelman syndrome with diabetes and kidney stones: A case report.

Medicine (Baltimore)

January 2025

The Department of Clinical Laboratory, Zhejiang Hospital, Hangzhou, China.

Rationale: Gitelman syndrome (GS) is a rare hereditary electrolyte disorder caused by mutations in the SLC12A3 gene. There is limited literature on the role of hydrochlorothiazide (HCT) testing and the SLC12A3 single heterozygous mutation in the diagnosis and management of patients with GS. In addition, cases of GS with concomitant kidney stones are rare.

View Article and Find Full Text PDF

Gestational diabetes mellitus (GDM), a transient form of diabetes that resolves postpartum, is a major risk factor for type 2 diabetes (T2D) in women. While the progression from GDM to T2D is not fully understood, it involves both genetic and environmental components. By integrating clinical, metabolomic, and genome-wide association study (GWAS) data, we identified associations between decreased sphingolipid biosynthesis and future T2D, in part through the allele of the gene in Hispanic women shortly after a GDM pregnancy.

View Article and Find Full Text PDF

Objective: To evaluate the therapeutic effects of Kuanxiong Aerosol (KXA) on ischemic stroke with reperfusion and elucidate the underlying pharmacological mechanisms.

Methods: In vivo pharmacological effects on ischemic stroke with reperfusion was evaluated using the transient middle cerebral artery occlusion (t-MCAO) mice model. To evaluate short-term outcome, 30 mice were randomly divided into vehicle group (n=15) and KXA group (n=15).

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!