Background: Prompt and accurate diagnosis is needed to prevent the untoward effects of anaemia on children. Although haematology analyzers are the gold standard for accurate measurement of haemoglobin or haematocrit for anaemia diagnosis, they are often out of the reach of most health facilities in resource-poor settings thus creating a care gap. We conducted this study to examine the agreement between a point-of-care device and haematology analyzer in determining the haematocrit levels in children and to determine its usefulness in diagnosing anaemia in resource-poor settings.
Methods: EDTA blood samples collected from participants were processed to estimate their haematocrits using the two devices (Mindray BC-3600 haematology analyzer and Portable Mission Hb/Haemotocrit testing system). A pairwise t-test was used to compare the haematocrit (PCV) results from the automated haematology analyzer and the portable haematocrit meter. The agreement between the two sets of measurements was assessed using the Bland and Altman method where the mean, standard deviation and limit of agreement of paired results were calculated.
Results: The intraclass and concordance correlation coefficients were 0.966 and 0.936. Sensitivity and specificity were 97.85% and 94.51% respectively while the positive predictive and negative predictive values were 94.79% and 97.73%. The Bland and Altman`s limit of agreement was -5.5-5.1 with the mean difference being -0.20 and a non-ignificant variability between the two measurements (p = 0.506).
Conclusion: Haematocrit determined by the portable testing system is comparable to that determined by the haematology analyzer. We therefore recommend its use as a point-of-care device for determining haematocrit in resource-poor settings where haematology analyzers are not available.
Download full-text PDF |
Source |
---|---|
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC4913193 | PMC |
http://dx.doi.org/10.4314/ejhs.v26i3.8 | DOI Listing |
Clin Chem Lab Med
January 2025
Department of Nephrology, Ghent University Hospital Ghent, Belgium.
Objectives: We evaluated the performance of a novel flow cell morphology analyzer AUTION EYE AI-4510 for counting particles in urine.
Methods: Analytical performance was assessed according to the EFLM European Urinalysis Guideline 2023. Trueness was compared by analyzing 1.
Open Forum Infect Dis
January 2025
CHU d'Orléans, Orléans, France.
Background: To better understand factors associated with virologic response, we retrospectively characterized the HIV proviruses of 7 people with HIV who received long-acting cabotegravir/rilpivirine (CAB/RPV-LA) and were selected according to the following criteria: virologic control achieved despite a history of viral replication on 1 or both corresponding antiretroviral classes (n = 6) and virologic failure (VF) after CAB/RPV-LA initiation (n = 1).
Methods: Last available blood samples before the initiation of CAB/RPV-LA were analyzed retrospectively. Near full-length HIV DNA genome haplotypes were inferred from Nanopore sequencing by the in vivo Genome Diversity Analyzer to search for archived drug resistance mutations (DRMs) and evaluate the frequency and intactness of proviruses harboring DRMs.
Artif Organs
January 2025
Division of Life Science and Medicine, School of Biomedical Engineering (Suzhou), University of Science and Technology of China, Hefei, China.
Background: Membrane oxygenators facilitate extracorporeal gas exchange, necessitating the monitoring of blood gas. Recent advances in normothermic machine perfusion (NMP) for ex vivo liver offer solutions to the shortage of donor liver. However, maintaining physiological blood gas levels during prolonged NMP is complex and costly.
View Article and Find Full Text PDFCurr Environ Health Rep
January 2025
School of Health Sciences, Purdue University, West-Lafayette, IN, 47906, USA.
Purpose Of Review: This review explores the use of Inductively Coupled Plasma Mass Spectrometry (ICP-MS) and X-ray Fluorescence (XRF) for quantifying metals and metalloids in biological matrices such as hair, nails, blood, bone, and tissue. It provides a comprehensive overview of these methodologies, detailing their technological limitations, application scopes, and practical considerations for selection in both laboratory and field settings. By examining traditional and novel aspects of each method, this review aims to guide researchers and clinical practitioners in choosing the most suitable analytical tool based on their specific needs for sensitivity, precision, speed, and sample preparation.
View Article and Find Full Text PDFScand J Clin Lab Invest
January 2025
Department of Medical Biochemistry, Health Science University, Istanbul Basaksehir Cam and Sakura City Hospital, Istanbul, Turkey.
This study assessed the reliability of Roche Accu-Chek Inform II glucometers in a real-world setting. A retrospective analysis was conducted on 6,695 paired results. Capillary samples were tested using Roche Accu-Chek Inform II glucometers, while venous samples were analyzed using Roche Cobas c503/702 analyzers.
View Article and Find Full Text PDFEnter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!