Background: Despite technological advances, the accuracy of continuous glucose monitoring (CGM) systems may not always be satisfactory with rapidly changing glucose levels, as is notable during exercise. We compare the performance of two current and widely used CGM systems, Dexcom G4 Platinum (Dexcom) and Medtronic Paradigm Veo Enlite system (Enlite), during both rest and exercise in adults with type 1 diabetes (T1D).
Research Design And Methods: Paired sensor and plasma glucose (PG) values (total of 431 data pairs for Dexcom and 425 for Enlite) were collected from 17 adults (37.3 ± 13.6 years) with T1D. To evaluate and compare the accuracy of sensor readings, criteria involving sensor bias (sensor minus PG levels), absolute relative difference (ARD), and percentage of readings meeting International Organization for Standardization (ISO) criteria were considered.
Results: Both Dexcom and Enlite performed equally well during the rest period, with respective mean/median biases of -0.12/-0.02 mmol/L versus -0.18/-0.40 (P = 0.78, P = 0.66) mmol/L and ARDs of 13.77/13.34% versus 12.38/11.95% (P = 0.53, P = 0.70). During exercise, sensor bias means/medians were -0.40/-0.21 mmol versus -0.26/-0.24 mmol/L (P = 0.67, P = 0.62) and ARDs were 22.53/15.13% versus 20.44/14.11% (P = 0.58, P = 0.68) for Dexcom and Enlite, respectively. Both sensors demonstrated significantly lower performance during exercise; median ARD comparison at rest versus exercise for both Dexcom and Enlite showed a P = 0.02. More data pairs met the ISO criteria for Dexcom and Enlite at rest, 73.6% and 76.9% compared with exercise 48.2% and 53.9%.
Conclusion: Dexcom and Enlite demonstrated comparable overall performances during rest and physical activity. However, a lower accuracy was observed during exercise for both sensors, necessitating a fine-tuning of their performance with physical activity.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1089/dia.2015.0394 | DOI Listing |
Pediatr Diabetes
March 2021
Division of Endocrinology and Diabetology, Department of Internal Medicine, Medical University of Graz, Austria.
The aim of this study was to assess accuracy of the three most commonly used continuous glucose monitoring (CGM) systems in almost real-life situation during a diabetes camp in children with type 1 diabetes (T1D) aged 9-14 years. Data was gathered during a 2-week summer camp under physicians' supervision. Out of 38 participating children with T1D (aged: 11.
View Article and Find Full Text PDFPediatr Diabetes
March 2021
Division of Endocrinology and Metabolism, Children's Hospital of Eastern Ontario, Ottawa, Canada.
To determine if pump therapy with continuous glucose monitoring offering low glucose suspend (LGS) decreases fear of hypoglycemia among children with type 1 diabetes and their parents. The CGM TIME trial is a multicenter randomized controlled trial that enrolled 144 children with type 1 diabetes for at least 1 year (mean duration 3.4 ± 3.
View Article and Find Full Text PDFDiabetes Technol Ther
October 2019
Department of Conservative Dentistry and Periodontology, University Hospital, Ludwig-Maximilian University, Munich, Germany.
Along with increased usage of continuous glucose monitors, flash glucose monitors, and patch pumps by patients with diabetes, the frequency of skin reactions has also increased. Skin irritation and itching can be annoying to users. However, more serious contact allergies to one or more components of the adhesives or plastic material of the housing of the devices can become lifelong.
View Article and Find Full Text PDFDiabetes Obes Metab
April 2019
Division of Endocrinology and Diabetology, Department of Internal Medicine, Medical University of Graz, Graz, Austria.
To compare the performance of a professional continuous glucose monitoring (proCGM) and a personal continuous glucose monitoring (persCGM) system worn in parallel under standardized conditions in individuals with type 1 diabetes (T1D), two CGM systems (iPro2 - proCGM; Minimed 640G - persCGM) worn in parallel using the same sensor (Enlite 2) were compared. Ten people with T1D were included in this single-centre, open-label study in which CGM performance was evaluated. The study consisted of a 24-hours inpatient phase (meals, exercise, glycaemic challenges) and a 4-day home phase.
View Article and Find Full Text PDFSensors (Basel)
June 2017
Institut d'Informàtica i Aplicacions, Universitat de Girona, Campus de Montilivi, s/n, Edifici P4, 17071 Girona, Spain.
Continuous glucose monitors (CGMs) are prone to inaccuracy due to time lags, sensor drift, calibration errors, and measurement noise. The aim of this study is to derive the model of the error of the second generation Medtronic Paradigm Veo Enlite (ENL) sensor and compare it with the Dexcom SEVEN PLUS (7P), G4 PLATINUM (G4P), and advanced G4 for Artificial Pancreas studies (G4AP) systems. An enhanced methodology to a previously employed technique was utilized to dissect the sensor error into several components.
View Article and Find Full Text PDFEnter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!