A series of measurements have been recently conducted to determine the cosmic-muon intensities and attenuation factors at various indoor and underground locations for a gamma spectrometer. For this purpose, a digital coincidence spectrometer was developed by using two BC408 plastic scintillation detectors and an XIA LLC Digital Gamma Finder (DGF)/Pixie-4 software and card package. The results indicate that the overburden in the building at surface level absorbs a large part of cosmic ray protons while attenuating the cosmic-muon intensity by 20-50%. The underground facility has the largest overburden of 39 m water equivalent, where the cosmic-muon intensity is reduced by a factor of 6. The study provides a cosmic-muon intensity measurement and overburden assessment, which are important parameters for analysing the background of an HPGe counting system, or for comparing the background of similar systems.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.jenvrad.2016.06.011DOI Listing

Publication Analysis

Top Keywords

cosmic-muon intensity
16
intensity measurement
8
measurement overburden
8
building surface
8
surface level
8
underground facility
8
scintillation detectors
8
counting system
8
cosmic-muon
5
overburden
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?