Impact of imaging approach on radiation dose and associated cancer risk in children undergoing cardiac catheterization.

Catheter Cardiovasc Interv

Division of Pediatric Cardiology, Department of Pediatrics (K.D.H., J.S.L, G.A.F.) and the Department of Radiology (C.W., T.Y., N.J., G.N.), Duke University Medical Center, Durham, North Carolina, and the Duke Clinical Research Institute (K.D.H., J.S.L.), Durham, North Carolina and the Department of Medicine, Division of Cardiology, and Department of Radiology, Columbia University Medical Center and New York-Presbyterian Hospital, New York NY (A.J.E.).

Published: April 2017

Objectives: To quantify the impact of image optimization on absorbed radiation dose and associated risk in children undergoing cardiac catheterization.

Background: Various imaging and fluoroscopy system technical parameters including camera magnification, source-to-image distance, collimation, antiscatter grids, beam quality, and pulse rates, all affect radiation dose but have not been well studied in younger children.

Methods: We used anthropomorphic phantoms (ages: newborn and 5 years old) to measure surface radiation exposure from various imaging approaches and estimated absorbed organ doses and effective doses (ED) using Monte Carlo simulations. Models developed in the National Academies' Biological Effects of Ionizing Radiation VII report were used to compare an imaging protocol optimized for dose reduction versus suboptimal imaging (+20 cm source-to-image-distance, +1 magnification setting, no collimation) on lifetime attributable risk (LAR) of cancer.

Results: For the newborn and 5-year-old phantoms, respectively ED changes were as follows: +157% and +232% for an increase from 6-inch to 10-inch camera magnification; +61% and +59% for a 20 cm increase in source-to-image-distance; -42% and -48% with addition of 1-inch periphery collimation; -31% and -46% with removal of the antiscatter grid. Compared with an optimized protocol, suboptimal imaging increased ED by 2.75-fold (newborn) and fourfold (5 years old). Estimated cancer LAR from 30-min of posteroanterior fluoroscopy using optimized versus suboptimal imaging, respectively was 0.42% versus 1.23% (newborn female), 0.20% versus 0.53% (newborn male), 0.47% versus 1.70% (5-year-old female) and 0.16% versus 0.69% (5-year-old male).

Conclusions: Radiation-related risks to children undergoing cardiac catheterization can be substantial but are markedly reduced with an optimized imaging approach. © 2016 Wiley Periodicals, Inc.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC5164876PMC
http://dx.doi.org/10.1002/ccd.26630DOI Listing

Publication Analysis

Top Keywords

radiation dose
12
children undergoing
12
undergoing cardiac
12
suboptimal imaging
12
imaging approach
8
dose associated
8
risk children
8
cardiac catheterization
8
camera magnification
8
versus suboptimal
8

Similar Publications

A combination of gold nanoparticles and laser photobiomodulation to boost antioxidant defenses in the recovery of muscle injuries caused by Bothrops jararaca venom.

Lasers Med Sci

January 2025

Laboratory of Pathophysiology Experimental, Postgraduate Program in Health Sciences, Universidade do Extremo Sul Catarinense (UNESC), Criciúma, SC, Brazil.

Unlabelled: This study aimed to evaluate gold nanoparticles (GNPs) and photobiomodulation (PBM), associated with antibothropic serum (AS), to treat a muscle lesion induced by Bothrops jararaca venom.

Methods: 108 Swiss male mice were used, divided into nine groups (n = 12) with different combinations of treatments. Animals were inoculated with 250 µg of B.

View Article and Find Full Text PDF

Three-Dimensional CT for Preprocedural Planning of PCI for Ostial Right Coronary Artery Lesions: A Randomized Controlled Pilot Trial.

Circ Cardiovasc Interv

January 2025

Department of Cardiology, Ziekenhuis Oost-Limburg, Genk, Belgium (D.M.F.v.d.B., E.M.P., E.W., D.C., E.M., B.F., M.V., J.D., K.A.).

Background: Geographic stent-ostium mismatch is an important predictor of target lesion failure after percutaneous coronary intervention of an aorto-ostial right coronary artery lesion. Optimal visualization of the aorto-ostial plane is crucial for precise stent implantation at the level of the ostium. This study investigates whether preprocedural 3-dimensional computed tomography (3DCT), with determination of the optimal viewing angle, would allow for more precise stent implantation and reduce procedure time, contrast, and radiation dose.

View Article and Find Full Text PDF

Dosimetric Planning Comparison for Left Ventricle Avoidance in Non-small Cell Lung Cancer Radiotherapy.

Cureus

December 2024

Physics and Engineering, London Regional Cancer Program, London, CAN.

Introduction: Radiation may unintentionally injure myocardial tissue, potentially leading to radiation-induced cardiac disease (RICD), with the net benefit of non-small cell lung cancer (NSCLC) radiotherapy (RT) due to the proximity of the lung and heart. RTOG-0617 showed a greater reduction in overall survival (OS) comparing higher doses to standard radiation doses in NSCLC RT. VHeart has been reported as an OS predictor in the first- and fifth-year follow-ups.

View Article and Find Full Text PDF

Background: Different doses of radiotherapy (RT) exert diverse effects on tumor immunity, although the precise irradiation method remains unknown. This study sought to elucidate the influence of combining different doses of RT with immune checkpoint inhibitors (ICIs) on the infiltration of CD8T cells within tumors, thereby augmenting the anti-tumor response.

Methods: Constructing a mouse model featuring bilateral lung cancer tumors subjected to high and low dose irradiation, the analysis of RNA transcriptome sequencing data and immunohistochemical validation for tumors exposed to various dosages guided the selection of the optimal low-dose irradiation scheme.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!