This paper presents a statistical evaluation of the responsivity data on a number of heat-flux sensors, calibrated using an electrical substitution radiometer as a transfer standard up to 5 W·cm(-2). The sensors, furnished by the customers, were of circular-foil or thermopile type. Comparison of the NIST and the customer measured responsivity values showed that the measurements agree within 3 % for more than half the number of sensors tested, so far. Considering the variation in the customer calibration techniques and the wide measuring range of the sensors used in the calibration, the agreement is encouraging.