Physiological vortices in the sinuses of Valsalva: An in vitro approach for bio-prosthetic valves.

J Biomech

UCL Cardiovascular Engineering Laboratory, UCL Mechanical Engineering, University College London, UK. Electronic address:

Published: September 2016

Purpose: The physiological flow dynamics within the Valsalva sinuses, in terms of global and local parameters, are still not fully understood. This study attempts to identify the physiological conditions as closely as possible, and to give an explanation of the different and sometime contradictory results in literature.

Methods: An in vitro approach was implemented for testing porcine bio-prosthetic valves operating within different aortic root configurations. All tests were performed on a pulse duplicator, under physiological pressure and flow conditions. The fluid dynamics established in the various cases were analysed by means of 2D Particle Image Velocimetry, and related with the achieved hydrodynamic performance.

Results: Each configuration is associated with substantially different flow dynamics, which significantly affects the valve performance. The configuration most closely replicating healthy native anatomy was characterised by the best hemodynamic performance, and any mismatch in size and position between the valve and the root produced substantial modification of the fluid dynamics downstream of the valve, hindering the hydrodynamic performance of the system. The worst conditions were observed for a configuration characterised by the total absence of the Valsalva sinuses.

Conclusion: This study provides an explanation for the different vortical structures described in the literature downstream of bioprosthetic valves, enlightening the experimental complications in valve testing. Most importantly, the results clearly identify the fluid mechanisms promoted by the Valsalva sinuses to enhance the ejection and closing phases, and this study exposes the importance of an optimal integration of the valve and root, to operate as a single system.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC5061069PMC
http://dx.doi.org/10.1016/j.jbiomech.2016.05.027DOI Listing

Publication Analysis

Top Keywords

vitro approach
8
bio-prosthetic valves
8
flow dynamics
8
valsalva sinuses
8
fluid dynamics
8
valve root
8
valve
5
physiological
4
physiological vortices
4
vortices sinuses
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!