Accuracy of transcutaneous bilirubin measurement in newborns after phototherapy.

J Perinatol

Department of Neonatology, Children Faculty Hospital Kosice and Medical Faculty, Pavol Jozef Safarik University Kosice, Slovakia.

Published: October 2016

Objective: We tested the accuracy of transcutaneous bilirubin (TcB) measure in newborns undergoing phototherapy.

Study Design: In a prospective study of 150 term Caucasian neonates, 255 measurements of total serum bilirubin (TSB) and TcB concentration were obtained 2 h after discontinuing phototherapy. TcB measurements obtained on the forehead, sternum, abdomen and covered lower abdomen were statistically compared with the corresponding TSB.

Result: TcB consistently underestimated TSB levels significantly. The smallest but significant difference between TSB and TcB was found on the lower abdomen. The correlation between TSB and TcB was found to be moderately close (r=0.4 to 0.5). TcB measurements were inaccurate, regardless of phototherapy technique (Bilibed, conventional phototherapy).

Conclusion: Phototherapy significantly interferes with the accuracy of transcutaneous bilirubinometry. TcB measurements performed 2 h after stopping phototherapy were not reliable, even if they were carried out on the unexposed body area. TSB assessment remains necessary, if treatment of hyperbilirubinemia is being considered.

Download full-text PDF

Source
http://dx.doi.org/10.1038/jp.2016.91DOI Listing

Publication Analysis

Top Keywords

accuracy transcutaneous
12
tsb tcb
12
tcb measurements
12
transcutaneous bilirubin
8
tcb
8
lower abdomen
8
phototherapy
5
tsb
5
bilirubin measurement
4
measurement newborns
4

Similar Publications

A Wearable Prototype Measuring PtcCO and SpO.

IEEE Biomed Circuits Syst Conf

October 2024

Department of Electrical and Computer Engineering, Worcester Polytechnic Institute, Worcester, MA 01609 USA.

The proper functioning of the respiratory system is evaluated by monitoring the exchange of blood oxygen and carbon dioxide. While wearable devices for monitoring both blood oxygen and carbon dioxide are emerging, wearable carbon dioxide monitors remain relatively rare. This paper introduces a novel wearable prototype that integrates the measurement of transcutaneous carbon dioxide and peripheral blood oxygen saturation on a miniaturized custom-designed printed circuit board.

View Article and Find Full Text PDF

This study develops biomimetic strategies for slip prevention in prosthetic hand grasps. The biomimetic system is driven by a novel slip sensor, followed by slip perception and preventive control. Here, we show that biologically inspired sensorimotor pathways can be restored between the prosthetic hand and users.

View Article and Find Full Text PDF

In vivo Raman spectroscopy for non-invasive transcutaneous glucose monitoring on animal models and human subjects.

Spectrochim Acta A Mol Biomol Spectrosc

December 2024

Institute of Photonics and Photon-Technology, Northwest University, Xi'an, Shaanxi 710127, China. Electronic address:

Non-invasive glucose monitoring represents a significant advancement in diabetes management and treatment as non-painful alternatives than finger-sticks tests. After developing an integrated Raman spectral system with a 785 nm laser, this study systematically explores the application of in vivo Raman spectroscopy for quantitative, noninvasive glucose monitoring. In addition to observing characteristic glucose spectral information from a mouse model, a strong spectral correlation was also recognized with the blood glucose concentration.

View Article and Find Full Text PDF

Neuromodulation techniques effectively intervene in cognitive function, holding considerable scientific and practical value in fields such as aerospace, medicine, life sciences, and brain research. These techniques utilize electrical stimulation to directly or indirectly target specific brain regions, modulating neural activity and influencing broader brain networks, thereby regulating cognitive function. Regulating cognitive function involves an understanding of aspects such as perception, learning and memory, attention, spatial cognition, and physical function.

View Article and Find Full Text PDF

Study Objectives: Clinical utility of home polysomnography in children with neuromuscular disorders is limited by lack of evidence that sleep-disordered breathing can be reliably identified and inability to diagnose hypoventilation as carbon dioxide is not measured.

Methods: This study aimed to determine feasibility, accuracy and parent satisfaction for home polysomnography performed with a Type 2 portable monitoring device and a transcutaneous CO monitor. Results of laboratory and home polysomnography were compared with Mann-Whitney U tests and random intercept regression models.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!