Estimation of Doppler shift frequency using selected phase information for high frame rate color flow mapping.

J Med Ultrason (2001)

Precision and Intelligence Laboratory, Tokyo Institute of Technology, 4259 Nagatsuta, Midori-ku, Yokohama, 226-8503, Japan.

Published: March 2004

Purpose: We describe a new approach to processing signals used to estimate the Doppler shift frequency in high frame-rate color flow mapping with fewer pulse transmissions. When an ultrasound pulse is transmitted to a large number of scatterers, the echoes from the scatterers overlap and interfere with one another. This interference causes the phase of the received echo signal to fluctuate, thus disturbing the estimated shift in Doppler frequency. The technique proposed here eliminates this disturbed phase information, leaving the remaining information for use in estimating the shift in Doppler frequency. The instantaneous frequency of the echo signal can serve as an index of the influence of interference.

Methods: To test this technique in vivo we used radio-frequency echo signals from the carotid artery for simulation and evaluated the error of the estimated Doppler shift frequency in several cases.

Conclusion: Performance was enhanced when the number of pulses transmitted was limited and this technique was used.

Download full-text PDF

Source
http://dx.doi.org/10.1007/s10396-003-0005-4DOI Listing

Publication Analysis

Top Keywords

doppler shift
12
shift frequency
12
color flow
8
flow mapping
8
echo signal
8
shift doppler
8
doppler frequency
8
frequency
6
shift
5
estimation doppler
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!