Maximum likelihood positioning algorithm for high-resolution PET scanners.

Med Phys

Department for Physics of Molecular Imaging Systems, Institute for Experimental Molecular Imaging, RWTH Aachen University, Aachen NRW 52074, Germany and Philips Research Europe, Aachen NRW 52074, Germany.

Published: June 2016

Purpose: In high-resolution positron emission tomography (PET), lightsharing elements are incorporated into typical detector stacks to read out scintillator arrays in which one scintillator element (crystal) is smaller than the size of the readout channel. In order to identify the hit crystal by means of the measured light distribution, a positioning algorithm is required. One commonly applied positioning algorithm uses the center of gravity (COG) of the measured light distribution. The COG algorithm is limited in spatial resolution by noise and intercrystal Compton scatter. The purpose of this work is to develop a positioning algorithm which overcomes this limitation.

Methods: The authors present a maximum likelihood (ML) algorithm which compares a set of expected light distributions given by probability density functions (PDFs) with the measured light distribution. Instead of modeling the PDFs by using an analytical model, the PDFs of the proposed ML algorithm are generated assuming a single-gamma-interaction model from measured data. The algorithm was evaluated with a hot-rod phantom measurement acquired with the preclinical hyperion II (D) PET scanner. In order to assess the performance with respect to sensitivity, energy resolution, and image quality, the ML algorithm was compared to a COG algorithm which calculates the COG from a restricted set of channels. The authors studied the energy resolution of the ML and the COG algorithm regarding incomplete light distributions (missing channel information caused by detector dead time). Furthermore, the authors investigated the effects of using a filter based on the likelihood values on sensitivity, energy resolution, and image quality.

Results: A sensitivity gain of up to 19% was demonstrated in comparison to the COG algorithm for the selected operation parameters. Energy resolution and image quality were on a similar level for both algorithms. Additionally, the authors demonstrated that the performance of the ML algorithm is less prone to missing channel information. A likelihood filter visually improved the image quality, i.e., the peak-to-valley increased up to a factor of 3 for 2-mm-diameter phantom rods by rejecting 87% of the coincidences. A relative improvement of the energy resolution of up to 12.8% was also measured rejecting 91% of the coincidences.

Conclusions: The developed ML algorithm increases the sensitivity by correctly handling missing channel information without influencing energy resolution or image quality. Furthermore, the authors showed that energy resolution and image quality can be improved substantially by rejecting events that do not comply well with the single-gamma-interaction model, such as Compton-scattered events.

Download full-text PDF

Source
http://dx.doi.org/10.1118/1.4950719DOI Listing

Publication Analysis

Top Keywords

energy resolution
28
resolution image
20
image quality
20
positioning algorithm
16
cog algorithm
16
algorithm
14
measured light
12
light distribution
12
missing channel
12
maximum likelihood
8

Similar Publications

High-resolution fluorescence imaging of ultracold atoms and molecules is paramount to performing quantum simulation and computation in optical lattices and tweezers. Imaging durations in these experiments typically range from a millisecond to a second, significantly limiting the cycle time. In this work, we present fast, 2.

View Article and Find Full Text PDF

The objective of this work was to explore the capabilities of a field emission gun scanning electron microscope (FEG-SEM) equipped with a transmission scanning electron detector (TSEM) and energy dispersive spectroscopy (EDS) to identify nanoscale chemical heterogeneities in a gas atomization reaction synthesis (GARS) steel sample. The results of this analysis were compared to the same study conducted with scanning transmission electron microscopy (STEM) with EDS mapping. TSEM-EDS was performed using the standard spectral analysis approach, i.

View Article and Find Full Text PDF

High-throughput and sustainable B vitamins analysis in nutritional supplements, vegetables, and fruits via 2D carbon microfiber fractionation system coupled with mass spectrometry.

Talanta

January 2025

College of Pharmacy, Yanbian University, Park Road 977, Yanji, 133002, Jilin Province, PR China; Department of Chemistry, Interdisciplinary Program of Biological Functional Molecules, College of Integration Science, Yanbian University, Park Road 977, Yanji, 133002, Jilin Province, PR China. Electronic address:

B vitamins are essential for energy metabolism, nervous health, blood production, and the immune system. Their quantification in nutritional supplements and food is mandatory to manage a correct daily intake and dosage. In this study, a fast and sustainable method for the analysis of 8 B vitamins (VB, VB, VB, nicotinamide, VB, VB, VB, VB) in real samples using a 2D-carbon microfiber fractionation system combined with a triple quadrupole mass spectrometer (2DμCFs-QqQ-MS/MS) is presented.

View Article and Find Full Text PDF

Adsorption structures and bonding states of cesium and barium adsorbed on various sites of vermiculite.

Sci Total Environ

January 2025

Department of Earth and Planetary Science, Graduate School of Science, The University of Tokyo, Bunkyo, Tokyo 113-0033, Japan; Isotope Science Center, The University of Tokyo, Bunkyo, Tokyo 113-0032, Japan. Electronic address:

The adsorption reaction on clay minerals is crucial for understanding the environmental behavior of various cations, including cesium (Cs). However, its details remain unclear because of multiple adsorption sites of the clay minerals, a significant difference between concentrations in the atomic-scale experiments and the actual environment, and difficulties of evaluating bonding states of the adsorbed cations. It is expected that systematic experiments at the atomic-scale with a wide concentration range and application of density functional theory (DFT) calculations overcome the problems and bring crucial insights to link laboratory experiment results with environmental sample analysis.

View Article and Find Full Text PDF

Continuous production of entropy and the corresponding energy dissipation is a defining characteristic of nonequilibrium systems. When a system's full chemical kinetic description is known, its entropy production rate can be computed from the microscopic rate constants. However, such a calculation typically underestimates energy dissipation when the states of the underlying system are mesoscopic, i.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!