Purpose: Metal artifact reduction (MAR) is a major problem and a challenging issue in x-ray computed tomography (CT) examinations. Iterative reconstruction from sinograms unaffected by metals shows promising potential in detail recovery. This reconstruction has been the subject of much research in recent years. However, conventional iterative reconstruction methods easily introduce new artifacts around metal implants because of incomplete data reconstruction and inconsistencies in practical data acquisition. Hence, this work aims at developing a method to suppress newly introduced artifacts and improve the image quality around metal implants for the iterative MAR scheme.

Methods: The proposed method consists of two steps based on the general iterative MAR framework. An uncorrected image is initially reconstructed, and the corresponding metal trace is obtained. The iterative reconstruction method is then used to reconstruct images from the unaffected sinogram. In the reconstruction step of this work, an iterative strategy utilizing unmatched projector/backprojector pairs is used. A ramp filter is introduced into the back-projection procedure to restrain the inconsistency components in low frequencies and generate more reliable images of the regions around metals. Furthermore, a constrained total variation (TV) minimization model is also incorporated to enhance efficiency. The proposed strategy is implemented based on an iterative FBP and an alternating direction minimization (ADM) scheme, respectively. The developed algorithms are referred to as "iFBP-TV" and "TV-FADM," respectively. Two projection-completion-based MAR methods and three iterative MAR methods are performed simultaneously for comparison.

Results: The proposed method performs reasonably on both simulation and real CT-scanned datasets. This approach could reduce streak metal artifacts effectively and avoid the mentioned effects in the vicinity of the metals. The improvements are evaluated by inspecting regions of interest and by comparing the root-mean-square errors, normalized mean absolute distance, and universal quality index metrics of the images. Both iFBP-TV and TV-FADM methods outperform other counterparts in all cases. Unlike the conventional iterative methods, the proposed strategy utilizing unmatched projector/backprojector pairs shows excellent performance in detail preservation and prevention of the introduction of new artifacts.

Conclusions: Qualitative and quantitative evaluations of experimental results indicate that the developed method outperforms classical MAR algorithms in suppressing streak artifacts and preserving the edge structural information of the object. In particular, structures lying close to metals can be gradually recovered because of the reduction of artifacts caused by inconsistency effects.

Download full-text PDF

Source
http://dx.doi.org/10.1118/1.4950722DOI Listing

Publication Analysis

Top Keywords

unmatched projector/backprojector
12
projector/backprojector pairs
12
iterative reconstruction
12
iterative mar
12
iterative
10
metal artifact
8
artifact reduction
8
x-ray computed
8
computed tomography
8
conventional iterative
8

Similar Publications

Counter examples for unmatched projector/backprojector in an iterative algorithm.

Chin J Acad Radiol

July 2019

Department of Engineering, Weber State University, 1447 Edvalson Street, Ogden, UT 84408, USA.

It is rather controversial whether it is justified to use an unmatched projector/backprojector pair in an iterative image reconstruction algorithm. One common concern of using an unmatched projector/backprojector pair is that the optimal solution cannot be reached. This concern is misleading and must be clarified.

View Article and Find Full Text PDF

Assessment of the impact of modeling axial compression on PET image reconstruction.

Med Phys

October 2017

Division of Imaging Sciences & Biomedical Engineering, King's College London, St Thomas' Hospital, London, SE1 7EH, UK.

Purpose: To comprehensively evaluate both the acceleration and image-quality impacts of axial compression and its degree of modeling in fully 3D PET image reconstruction.

Method: Despite being used since the very dawn of 3D PET reconstruction, there are still no extensive studies on the impact of axial compression and its degree of modeling during reconstruction on the end-point reconstructed image quality. In this work, an evaluation of the impact of axial compression on the image quality is performed by extensively simulating data with span values from 1 to 121.

View Article and Find Full Text PDF

Purpose: Metal artifact reduction (MAR) is a major problem and a challenging issue in x-ray computed tomography (CT) examinations. Iterative reconstruction from sinograms unaffected by metals shows promising potential in detail recovery. This reconstruction has been the subject of much research in recent years.

View Article and Find Full Text PDF

GPU-accelerated iterative reconstruction from Compton scattered data using a matched pair of conic projector and backprojector.

Comput Methods Programs Biomed

July 2016

Department of Electronic Engineering, Paichai University, 155-40 Baejae-Ro, Doma-Dong, Seo-Gu, Daejeon 35345, Republic of Korea. Electronic address:

Background And Objective: Iterative reconstruction from Compton scattered data is known to be computationally more challenging than that from conventional line-projection based emission data in that the gamma rays that undergo Compton scattering are modeled as conic projections rather than line projections. In conventional tomographic reconstruction, to parallelize the projection and backprojection operations using the graphics processing unit (GPU), approximated methods that use an unmatched pair of ray-tracing forward projector and voxel-driven backprojector have been widely used. In this work, we propose a new GPU-accelerated method for Compton camera reconstruction which is more accurate by using exactly matched pair of projector and backprojector.

View Article and Find Full Text PDF

Unmatched projector/backprojector pairs in an iterative reconstruction algorithm.

IEEE Trans Med Imaging

May 2000

Department of Radiology, University of Utah, Salt Lake City 84108-1218, USA.

Computational burden is a major concern when an iterative algorithm is used to reconstruct a three-dimensional (3-D) image with attenuation, detector response, and scatter corrections. Most of the computation time is spent executing the projector and backprojector of an iterative algorithm. Usually, the projector and the backprojector are transposed operators of each other.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!