Purpose: To compare the size of ablation zones derived from nonperfused ex vivo livers with ablation zones created using an ex vivo perfused porcine liver model.

Materials And Methods: Six fresh porcine livers were used to evaluate microwave ablation (MWA). Perfused (n = 3) and nonperfused (n = 3) livers were warmed to 37°C by oxygenated, O-positive human blood reconstituted with Ringer solution, using an organ perfusion circuit. During MWA, perfusion was discontinued in the nonperfused group and maintained in the perfused group. After MWA (140 watts × 2 min at 2.45 GHz) with the Acculis MTA System (AngioDynamics, Latham, New York), ablation zones were bisected sagittally. Sections were stained with nicotinamide adenine dinucleotide (NADH) and hematoxylin-eosin to assess viability of cells in ablation and marginal zones.

Results: Comparison of 22 MWA zones (9 in perfused group, 13 in nonperfused group) was performed. Ablation zones demonstrated a central "white" and peripheral "red" zone. Cells in the white zone were nonviable with no NADH staining. The red zone showed progressive NADH staining toward the periphery, suggesting incomplete cell death. White and red zones of the perfused group were significantly smaller compared with the nonperfused group (short axis, 17.8 mm ± 2.7 vs 21.1 mm ± 3.2, P = .003; long axis, 40.69 mm ± 3.9 vs 39.63 mm ± 5.2, P = .44; intermediate zone,1.33 mm ± 0.04 vs 2.7 mm ± 0.14, P < .0001; mean ± SD).

Conclusions: MWA algorithms provided by this manufacturer are based on nonperfused organ data, which overestimate ablation zone size. Data from perfused liver models may be required for more accurate dosimetry guidelines.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.jvir.2016.03.006DOI Listing

Publication Analysis

Top Keywords

ablation zones
20
nonperfused group
12
perfused group
12
ablation
8
microwave ablation
8
porcine liver
8
zones perfused
8
nadh staining
8
zones
7
nonperfused
6

Similar Publications

Objective: This study aimed to introduce and evaluate a novel software-based system, BioTrace, designed for real-time monitoring of thermal ablation tissue damage during image-guided radiofrequency ablation for hepatocellular carcinoma (HCC).

Methods: BioTrace utilizes a proprietary algorithm to analyze the temporo-spatial behavior of thermal gas bubble activity during ablation, as seen in conventional B-mode ultrasound imaging. Its predictive accuracy was assessed by comparing the ablation zones it predicted with those annotated by radiologists using contrast-enhanced computed tomography (CECT) 24 hours post-treatment, considered the gold standard.

View Article and Find Full Text PDF

Ultrasound echo decorrelation imaging can successfully monitor and control thermal ablation of animal liver and tumor tissue ex vivo and in vivo. However, normal and diseased human liver has substantially different physical properties that affect echo decorrelation. Here, effects of human liver tissue condition on ablation guidance by three-dimensional echo decorrelation imaging are elucidated in experiments testing closed-loop control of radiofrequency ablation (RFA) in normal and diseased human liver tissue ex vivo.

View Article and Find Full Text PDF

Radiofrequency ablation (RFA) is a minimally invasive procedure that utilizes localized heat to treat tumors by inducing localized tissue thermal damage. The present study aimed to evaluate the temperature evolution and spatial distribution, ablation size, and reproducibility of ablation zones in ex vivo liver, kidney, and lung using a commercial device, i.e.

View Article and Find Full Text PDF

Purpose: Contrast-enhanced CT (CECT) may be performed immediately following microwave liver ablation for assessment of ablative margins. However, practices and protocols vary among institutions. Here, we compare a standardized bolus-tracked biphasic CECT protocol and compare this with a single venous phase fixed delay protocol for ablation zone (AZ) assessment.

View Article and Find Full Text PDF

Purpose: Pulsed electrical field (PEF) ablation is an energy-based technique used to treat a range of cancers by irreversible electroporation (IRE). Our objective was to use computational and plant-based models to characterize the electric field distribution and ablation zones induced with a commercial 8-needle array-based applicator intended for treatment of skin cancer when high-frequency IRE (H-FIRE) pulses are applied. Electric field characterisation of this device was not previously assessed.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!