Purpose: To present a metric for assessing the in vitro optical quality of rotationally symmetrical optical elements based on volume calculation under the surface defined by the axial modulation transfer function (MTF).

Setting: University of Valencia, Valencia, Spain.

Design: Experimental study.

Methods: The metric volume under the axial MTF was used to assess the optical quality of 2 rotationally symmetrical multifocal intraocular lenses (IOLs) within various defocus intervals (0.50 diopter [D], 0.75 D, and 1.00 D) and at various spatial frequency intervals (7.5 cycles per degree [cpd], 15.0 cpd, and 30.0 cpd).

Results: The far focus of the bifocal IOL yielded higher volume values at all spatial frequencies and defocus intervals than the trifocal IOL. The results for the near focus were similar for both IOLs. In addition, the trifocal IOL provided a distinct focus for intermediate vision.

Conclusions: The volume under the axial MTF proved to be a useful tool for objective evaluation of multifocal IOLs. Moreover, it can be applied to evaluate the optical quality of every rotationally symmetrical IOL or optical element.

Financial Disclosure: None of the authors has a financial or proprietary interest in any material or method mentioned.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.jcrs.2015.11.049DOI Listing

Publication Analysis

Top Keywords

optical quality
16
quality rotationally
12
rotationally symmetrical
12
assessing in vitro
8
in vitro optical
8
transfer function
8
volume axial
8
axial mtf
8
defocus intervals
8
trifocal iol
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!