A method has been developed for quantifying movement and wear of the acetabular component (cup) of total hip replacements (THR) from routine postoperative and review radiographs. The method uses both interactive and automatic computer image analysis techniques. Dimensions of the prosthesis are used to scale the measurements and so overcome variation in radiographic alignment. The application of the method is illustrated by retrospective investigations of cup migration and wear using review radiographs taken over a follow-up of at least 12 years.

Download full-text PDF

Source
http://dx.doi.org/10.1016/0141-5425(89)90153-2DOI Listing

Publication Analysis

Top Keywords

image analysis
8
review radiographs
8
prosthetic hip
4
hip failure
4
failure retrospective
4
retrospective radiograph
4
radiograph image
4
analysis acetabular
4
acetabular cup
4
cup method
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!