Purpose: Cardiac motion is a dominant source of physiological noise (PN) in myocardial arterial spin labeled (ASL) perfusion imaging. This study investigates the sensitivity to heart rate variation (HRV) of double-gated myocardial ASL compared with the more widely used single-gated method.

Methods: Double-gating and single-gating were performed on 10 healthy volunteers (n = 10, 3F/7M; age, 23-34 years) and eight heart transplant recipients (n = 8, 1F/7M; age, 26-76 years) at rest in the randomized order. Myocardial blood flow (MBF), PN, temporal signal-to-noise ratio (SNR), and HRV were measured.

Results: HRV ranged from 0.2 to 7.8 bpm. Double-gating PN did not depend on HRV, while single-gating PN increased with HRV. Over all subjects, double-gating provided a significant reduction in global PN (from 0.20 ± 0.15 to 0.11 ± 0.03 mL/g/min; P = 0.01) and per-segment PN (from 0.33 ± 0.23 to 0.21 ± 0.12 mL/g/min; P < 0.001), with significant increases in global temporal SNR (from 11 ± 8 to 18 ± 8; P = 0.02) and per-segment temporal SNR (from 7 ± 4 to 11 ± 12; P < 0.001) without significant difference in measured MBF.

Conclusion: Single-gated myocardial ASL suffers from reduced temporal SNR, while double-gated myocardial ASL provides consistent temporal SNR independent of HRV. Magn Reson Med 77:1975-1980, 2017. © 2016 International Society for Magnetic Resonance in Medicine.

Download full-text PDF

Source
http://dx.doi.org/10.1002/mrm.26282DOI Listing

Publication Analysis

Top Keywords

double-gated myocardial
8
myocardial asl
8
asl perfusion
8
perfusion imaging
8
heart rate
8
rate variation
8
hrv
5
imaging robust
4
robust heart
4
variation purpose
4

Similar Publications

Correction of respiratory and cardiac motion in cardiac PET/MR using MR-based motion modeling.

Phys Med Biol

November 2018

Translational and Molecular Imaging Institute, Icahn School of Medicine at Mount Sinai, One Gustave Levy Pl, New York, NY 10029, United States of America. Author to whom any correspondence should be addressed.

Cardiac positron emission tomography (PET) imaging suffers from image blurring due to the constant motion of the heart that can impact interpretation. Hybrid PET/magnetic resonance (MR) has the potential to use radiation-free MR imaging to correct for the effects of cardio-respiratory motion in the PET data, improving qualitative and quantitative PET imaging in the heart. The purpose of this study was (i) to implement a MR image-based motion-corrected PET/MR method and (ii) to perform a proof-of-concept study of quantitative myocardial PET data in patients.

View Article and Find Full Text PDF

Purpose: Cardiac motion is a dominant source of physiological noise (PN) in myocardial arterial spin labeled (ASL) perfusion imaging. This study investigates the sensitivity to heart rate variation (HRV) of double-gated myocardial ASL compared with the more widely used single-gated method.

Methods: Double-gating and single-gating were performed on 10 healthy volunteers (n = 10, 3F/7M; age, 23-34 years) and eight heart transplant recipients (n = 8, 1F/7M; age, 26-76 years) at rest in the randomized order.

View Article and Find Full Text PDF

Purpose: Cardiac imaging suffers from both respiratory and cardiac motion. One of the proposed solutions involves double gated acquisitions. Although such an approach may lead to both respiratory and cardiac motion compensation there are issues associated with (a) the combination of data from cardiac and respiratory motion bins, and (b) poor statistical quality images as a result of using only part of the acquired data.

View Article and Find Full Text PDF

Objectives: This study sought to determine whether arterial spin labeled (ASL) cardiac magnetic resonance (CMR) is capable of detecting clinically relevant increases in regional myocardial blood flow (MBF) with vasodilator stress testing in human myocardium.

Background: Measurements of regional myocardial perfusion at rest and during vasodilatation are used to determine perfusion reserve, which indicates the presence and distribution of myocardial ischemia. ASL CMR is a perfusion imaging technique that does not require any contrast agents, and is therefore safe for use in patients with end-stage renal disease, and capable of repeated or continuous measurement.

View Article and Find Full Text PDF

Purpose: Imaging of moving organs using the PET leads to blurred images due to long acquisition times. Simultaneous cardiac and respiratory gating of list-mode PET/CT is evaluated with the aim to improve image quality and assess the organ movement.

Methods: We performed a N-13 ammonia PET/CT scan with a human volunteer, using the Siemens Biograph Sensation 16 scanner with list-mode acquisition.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!