Preclinical metrics to predict through-focus visual acuity for pseudophakic patients.

Biomed Opt Express

AMO Groningen, van Swietenlaan, Groningen, 9728 NX, The Netherlands.

Published: May 2016

This study compares the clinical through-focus visual acuity (VA) in patients implanted with different intraocular lens (IOL) to optical bench testing of the same IOLs to evaluate the suitability of optical metrics of predicting clinical VA. Modulation transfer function and phase transfer function for different spatial frequencies and US Air Force pictures were measured using an optical bench for two monofocal IOLs, three multifocal IOLs and an extended range of vision IOL. Four preclinical metrics were calculated and compared to the clinical through-focus VA collected in three different clinical studies (243 patients in total). All metrics were well correlated (R(2)≥0.89) with clinical data and may be suitable for predicting through-focus VA in pseudophakic eyes.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC4871088PMC
http://dx.doi.org/10.1364/BOE.7.001877DOI Listing