Purpose: Pulmonary vein isolation with radiofrequency energy is widely used as a strategy for catheter ablation of atrial fibrillation (AF). Anatomically designed catheters have been developed to increase the efficiency of AF ablation procedures. The second-generation circular ablation catheter, PVAC GOLD, was re-designed to improve energy delivery and mitigate emboli. We investigated the procedural efficiency, biophysics, and chronic efficacy of PVAC GOLD in patients with AF.
Methods: We consecutively enrolled 40 patients (60 ± 11 years) with highly symptomatic, drug refractory AF. The first 20 patients were treated with the first-generation PVAC. The subsequent 20 patients were treated with the second-generation PVAC GOLD catheter. All patients were followed up at 3, 6, and 12 months.
Results: All 164 targeted PVs were successfully isolated. Ablations performed with PVAC GOLD showed a significant reduction in total number of ablations needed for PVI, fluoroscopy, and procedure times compared to PVAC (34.7 ± 7.0 vs. 27.0 ± 6.5; p = 0.009), fluoroscopy (29.5 ± 9.5 vs. 23.4 ± 7.0; p = 0.026), and procedure time (93.8 ± 18.9 vs. 83.1 ± 10.6; p = 0.033). PVAC GOLD showed improved biophysics including a reduction of low power ablations and an increase in mean effective energy delivery. At 12 months follow-up, AF recurrence rates were comparable in the two groups (35 vs. 30 %; p = 0.735). There were no adverse events.
Conclusions: The redesigned PVAC GOLD catheter demonstrates a reduction in radiofrequency ablation and procedure time and improved biophysics while maintaining chronic efficacy compared to the first-generation PVAC.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1007/s10840-016-0140-7 | DOI Listing |
Arq Bras Cardiol
June 2024
Unidade de Arritmia, Instituto do Coração (InCor), Hospital das Clínicas da Faculdade de Medicina da Universidade de São Paulo, São Paulo, SP - Brasil.
Background: There are no randomized studies comparing the maintenance of sinus rhythm after catheter ablation (CA) concerning treatment with antiarrhythmic drugs (AA) in elderly patients with paroxysmal atrial fibrillation (AF).
Objectives: To compare the clinical results of pulmonary vein (PV) isolation with the second-generation PVAC Gold catheter against AA treatment in elderly people with recurrent symptomatic paroxysmal AF, refractory to at least one AA, and without structural heart disease.
Methods: Sixty patients with paroxysmal AF ≥ 65 years old were randomized to two forms of treatment: group 1: CA and group 2: AA drugs.
J Interv Card Electrophysiol
November 2022
Emory University School of Medicine, Atlanta, GA, USA.
J Interv Card Electrophysiol
November 2022
Department of Cardiology, St. Antonius Hospital, Nieuwegein, The Netherlands.
Purpose: The second-generation multi-electrode catheter, PVAC Gold, was designed to improve the safe delivery of phased radiofrequency energy using a "single shot" approach for pulmonary vein isolation (PVI), while retaining efficacy. This large registry presents long-term performance in a daily practice setting.
Methods: A total of 1011 patients undergoing first time ablation for atrial fibrillation (AF) using PVAC Gold were included, 639 patients with PVI for paroxysmal AF (PAF PVI) and 372 patients with persistent or long-standing persistent AF, divided into 175 patients receiving PVI only (PersAF PVI) and 197 patients receiving PVI with additional substrate ablation (PersAF PVI +).
Zhongguo Yi Liao Qi Xie Za Zhi
September 2021
Zhejiang Medical Laboratory of Pulsed Power Technology, Hangzhou, 310003.
Pulsed electric field(PEF) provides high-energy instantaneous pulse and release energy to myocardial cell membrane, resulting in irreversible electroporation and causes myocardial cell contents leakage, destruction of intracellular homeostasis, cell death, and slight inflammatory response. PEF as non-thermal energy promotes the design and application of arrhythmia ablation catheter to enter a new stage. There are currently limited clinical studies that have proved the safety and effectieness of Farawave PEF catheter, PVAC GOLD PEF catheter, Lattice-tip Sphere-9 PEF and radiofrequency (RF) catheter used for atrial fibrillation ablation, but still need further discussion.
View Article and Find Full Text PDFEnter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!