The current study investigated the incidence of complications after surgery for distal radial fractures. This multicenter retrospective study was conducted at 11 institutions. A total of 824 patients who had distal radius fractures that were treated surgically between January 2010 and August 2012 were identified. The study patients were older than 18 years and were observed for at least 12 weeks after surgery for distal radius fractures with a volar locking plate. Sex, age, fracture type according to AO classification, implants, wrist range of motion, grip strength, fracture consolidation rate, and complications were studied. Analysis included 694 patients, including 529 women and 165 men, with a mean age of 64 years. The mean follow-up period was 27 weeks. The fracture consolidation rate was 100%. There were 52 complications (7.5%), including 18 cases of carpal tunnel syndrome, 12 cases of peripheral nerve palsy, 8 cases of trigger digit, 4 cases of tendon rupture (none of the flexor pollicis longus), and 10 others. There was no rupture of the flexor pollicis longus tendon because careful attention was paid to the relationship between the implant and the tendon. Peripheral nerve palsy may have been caused by intraoperative traction in 7 cases, temporary fixation by percutaneous Kirschner wires in 3 cases, and axillary nerve block in 1 case; 1 case appeared to be idiopathic. Tendon ruptures were mainly caused by mechanical stress. [Orthopedics.2016; 39(5):e893-e896.].
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.3928/01477447-20160517-05 | DOI Listing |
JBMR Plus
February 2025
Department of Medicine, School of Clinical Sciences at Monash Health, Monash University, Clayton, VIC 3168, Australia.
Quantifying precision error for DXA, peripheral QCT (pQCT), and HR-pQCT is crucial for monitoring longitudinal changes in body composition and musculoskeletal outcomes. Agreement and associations between bone variables assessed using pQCT and second-generation HR-pQCT are unclear. This study aimed to determine the precision of, and agreement and associations between, bone variables assessed via DXA, pQCT, and second-generation HR-pQCT.
View Article and Find Full Text PDFEur J Orthop Surg Traumatol
January 2025
Stony Brook University Hospital, Stony Brook, USA.
Purpose: Diabetes mellitus (DM) is a well-established risk factor for postoperative complications. Distal radius fractures (DRFs) are a common orthopedic injury and often require open reduction and internal fixation (ORIF). The rise of ORIF utilization warrants investigation into factors that may expose patients to postoperative complications following DRF ORIF.
View Article and Find Full Text PDFFront Bioeng Biotechnol
January 2025
Department of Orthopedics, Yangpu Hospital, School of Medicine, Tongji University, Shanghai, China.
Background: The most common postoperative complication of the Sauvé‒Kapandji (S-K) procedure is proximal ulnar stump instability. The distal oblique bundle (DOB) is a stable tissue used to stabilize the distal radioulnar joint. This study created finite-element models of the distal oblique bundle (DOB) to determine its effect on the proximal ulnar stump instability encountered during the Sauvé‒Kapandji procedure.
View Article and Find Full Text PDFOrthop Traumatol Surg Res
January 2025
Ankara University Faculty of Medicine, Orthopaedics and Traumatology Department, Hand Surgery Unit, Ankara, Turkey.
Introduction: Although there is no consensus in the literature, it is believed that the Soong classification system and fracture pattern are risk factors for plate removal in distal radius fractures.
Hypothesis: The aim of this large-scale study was to evaluate the relationship between Soong classification, fracture pattern, and implant removal in distal radius fractures.
Materials And Methods: We retrospectively evaluated 795 patients who underwent surgery using a volar locking plate for distal radius fractures at our clinic between 2005 and 2022.
Arch Osteoporos
January 2025
Department of Orthopaedics and Traumatology, Queen Mary Hospital The University of Hong Kong, 102 Pok Fu Lam Road, Hong Kong, China.
Unlabelled: Grip strength measurement, as a surrogate of sarcopenia diagnosis, effectively predicts secondary fracture risk in distal radius fracture patients. This simple tool enhances clinical practice by identifying high-risk patients for targeted interventions, potentially preventing or reversing functional decline and recurrent fractures.
Purpose: To evaluate grip strength and hand muscle cross-sectional area as predictors of secondary fracture risk in patients with a history of distal radius fracture (DRF), serving as surrogates of the diagnosis of sarcopenia.