In patients presenting non-high clinical pretest probability (PTP), a negative d-dimer can exclude venous thromboembolism without imaging tests. However, each d-dimer assay should be validated in prospective studies. We evaluated an automated d-dimer immunoassay using the Sclavo Auto d-dimer (Sclavo Diagnostics Int, Sovicille, Italy) provided by Dasit Diagnostica (Cornaredo, Milan, Italy). Three hundred two consecutive outpatients suspected of leg deep vein thrombosis (DVT) with non-high PTP were included. The Sclavo Auto d-dimer assay was evaluated on 2 analyzers (Sysmex CA-7000 and Sysmex CS-2100; Sysmex Corporation, Kobe, Japan, provided by Dasit). The cutoff value (200 ng/mL) was established a priori. Prevalence of DVT was 11.9%. Since no false-negative patients were detected, the sensitivity and negative predictive values (NPVs) were 100% (sensitivity = CA-7000: 100% [95% confidence interval, CI: 93.3-100], CS-2100: 100% [95% CI: 93.3-100]; NPV = CA-7000: 100% [95% CI: 97.9-100], CS-2100: 100% [95% CI: 98.0-100]). Specificity was 65.4% (95% CI: 59.4-71.1) and 69.2% (95% CI: 63.3-74.7) for CA-7000 and CS-2100, respectively. Specificity increased when a higher cutoff value (234 ng/mL) was used for patients aged ≥60 years without compromising the safety. Assay reproducibility was satisfactory at concentrations near the cutoff value (total coefficient of variations <10%). In conclusion, the Sclavo Auto d-dimer assay was accurate when used for DVT diagnostic workup in outpatients with non-high PTP. Based on its high sensitivity and NPV, it can be used as a stand-alone test in outpatients with non-high PTP. Given its high specificity, the number of patients in whom further imaging techniques can be avoided increased, improving the yield of the test.

Download full-text PDF

Source
http://dx.doi.org/10.1177/1076029616649436DOI Listing

Publication Analysis

Top Keywords

100% [95%
16
d-dimer assay
12
sclavo auto
12
auto d-dimer
12
deep vein
8
vein thrombosis
8
provided dasit
8
ca-7000 100%
8
cs-2100 100%
8
d-dimer
7

Similar Publications

Article Synopsis
  • The study measured fibrinogen fluorescence at temperatures between 20 and 80 degrees Celsius across different pH levels.
  • It was found that raising the temperature from 20 to 40 degrees Celsius did not change the structure of fibrinogen in solutions with pH between 4.5 and 9.3.
  • However, temperatures between 40 to 50 degrees Celsius caused some structural changes in neutral solutions, and temperatures above 50-55 degrees Celsius led to significant denaturation of the fibrinogen molecule.
View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!