Purpose. To determine the intraexaminer repeatability and interexaminer reproducibility of tear meniscus height (TMH) and noninvasive Keratograph tear breakup time (NIKBUT) measurements obtained with the Keratograph 5M (K5M) in a sample of healthy and dry eye populations. Methods. Forty-two patients with dry eye disease (DED group) and 42 healthy subjects (healthy group) were recruited in this prospective study. In all subjects, each eye received 3 consecutive measurements using the K5M for the TMH and NIKBUTs (NIKBUT-first and NIKBUT-average). And then a different examiner repeated the measurements. The repeatability and reproducibility of measurements were assessed by the coefficient of variation (CV) and intraclass correlation coefficient (ICC). Results. The repeatability and reproducibility of TMH and NIKBUTs were good in both DED and healthy groups (CV% ≤ 26.1% and ICC ≥ 0.75 for all measurements). Patients with DED showed better intraexaminer repeatability for NIKBUTs, but worse for TMH than healthy subjects. Average TMH, NIKBUT-first, and NIKBUT-average were significantly lower in DED group than in healthy group (all P values < 0.05). Conclusions. Measurements of TMH and NIKBUTs obtained with the K5M may provide a simple, noninvasive screening test for dry eye with acceptable repeatability and reproducibility. The NIKBUTs were more reliable, but TMH was less reliable in patients with DED.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC4844888PMC
http://dx.doi.org/10.1155/2016/8013621DOI Listing

Publication Analysis

Top Keywords

repeatability reproducibility
16
dry eye
16
tmh nikbuts
12
noninvasive keratograph
8
measurements patients
8
patients dry
8
eye disease
8
intraexaminer repeatability
8
ded group
8
group healthy
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!