The effect of time in use on the display performance of the iPad.

Br J Radiol

1 Centre for Online Health, The University of Queensland, Brisbane, QLD, Australia.

Published: July 2016

Objective: The aim of this study was to evaluate changes to the luminance, luminance uniformity and conformance to the digital imaging and communication in medicine greyscale standard display function (GSDF) as a function of time in use for the iPad.

Methods: Luminance measurements of the American Association of Physicists in Medicine (AAPM) Group 18 task group (TG18) luminance uniformity and luminance test patterns (TG18-UNL and TG18-LN8) were performed using a calibrated near-range luminance meter. Nine sets of measurements were taken, where the time in use of the iPad ranged from 0 to 2500 h.

Results: The maximum luminance (Lmax) of the display decreased (367-338 cdm(-2)) as a function of time. The minimum luminance remained constant. The maximum non-uniformity coefficient was 11%. Luminance uniformity decreased slightly as a function of time in use. The conformance of the iPad deviated from the GSDF curve at commencement of use. Deviation did not increase as a function of time in use.

Conclusion: This study has demonstrated that the iPad display exhibits luminance degradation typical of liquid crystal displays. The Lmax of the iPad fell below the American College of Radiology-AAPM-Society of Imaging Informatics in Medicine recommendations for primary displays (>350 cdm(-2)) at approximately 1000 h in use. The Lmax recommendation for secondary displays (>250 cdm(-2)) was exceeded during the entire study. The maximum non-uniformity coefficient did not exceed the recommendations for either primary or secondary displays. The deviation from the GSDF exceeded the recommendations of the TG18 for use as either a primary or secondary display.

Advances In Knowledge: The brightness, uniformity and contrast response are reasonably stable over the useful lifetime of the device; however, the device fails to meet the contrast response standard for either a primary or secondary display.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC5257299PMC
http://dx.doi.org/10.1259/bjr.20150657DOI Listing

Publication Analysis

Top Keywords

function time
16
luminance uniformity
12
primary secondary
12
luminance
10
maximum non-uniformity
8
non-uniformity coefficient
8
recommendations primary
8
secondary displays
8
contrast response
8
time
6

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!