Purpose: To determine conditions for the safe use of radiofrequency (RF) tissue ablation probes that avoid damaging suture material.
Methods: Four sutures made of 3 different materials commonly used in arthroscopic procedures were analyzed in a saline bath related to effects of RF-produced heat by proximity, duration, and intensity settings measuring burn-through time and ultimate load to failure. The parameters tested were electrode-to-suture distance, power setting, and the presence of tendon tissue or metallic anchor eyelets. Outcome variables were the burn-through time and the ultimate failure load of differently treated suture samples.
Results: Mean burn-through time for suture in direct contact with the RF probe ranged from 57.2 to 14.7 seconds for ultra-high-molecular-weight polyethylene (UHMWPE) sutures, 1.1 seconds for polydioxanone suture, and 0.8 seconds for polyethylene terephthalate suture. One of the UHMWPE sutures was capable of withstanding 3 seconds of direct contact with the RF probe without any compromise in tensile strength. No suture material tested had any mechanical change as long as the RF probe was kept 1 mm from the suture.
Conclusions: Heat from RF tissue ablation probes can cause undetected damage. High-strength UHMWPE sutures were less sensitive to an RF treatment than polyester sutures. The use of different test substrates did not significantly influence the burn through time.
Clinical Relevance: Heat from RF probes can damage some suture material if direct contact is made even briefly. The use of RF devices may be safe for the suture when a distance between probe and suture of >1 mm is maintained. Suture made from UHMWPE may tolerate up to 3 seconds of RF probe contact and not sustain significant damage. Surgeons must use great care when using RF devices in the vicinity of suture placement.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1016/j.arthro.2016.03.006 | DOI Listing |
J Craniofac Surg
September 2024
Department of Tuina, Sanming Hospital of Integrated Chinese and Western Medicine, Fujian University of Chinese Medicine, Sanming.
A deviated nasal septum leads to congestion and headaches. Surgery is the primary treatment, requiring careful postoperative septum positioning to prevent bleeding and hematoma. The study compared this method with nasal packing and traditional nasal septum suturing regarding surgical time, patient pain, nasal obstruction, and bleeding after septoplasty.
View Article and Find Full Text PDFAppl Biochem Biotechnol
January 2025
Department of Internal Medicine-Cardiovascular, Guangzhou Twelfth People's Hospital, No.1, Tianqiang Road, Tianhe District, Guangzhou City, Guangdong Province, 510620, China.
Myocardial infarction (MI) is a coronary artery-related disease that seriously threatens human life and is the leading cause of sudden death worldwide, where a lack of nutrients and oxygen leads to an inflammatory response and death of cardiomyocytes. Ferroptosis is a form of non-apoptotic cell death associated with metabolic dysfunction, resulting in abnormal breakdown of glutamine and iron-dependent accumulation of reactive oxygen species (ROS) during metabolism. However, the molecular mechanism of ferroptosis in the pathogenesis of MI and the function of Klotho and KRAS on ferroptosis during MI remain unclear.
View Article and Find Full Text PDFAnimal Model Exp Med
January 2025
Department of Orthopaedic Surgery, The 909th Hospital, School of Medicine, Xiamen University, Zhangzhou, China.
Backgroud: Intervertebral disc degeneration (IDD) is one of the common degenerative diseases. Due to ethical constraints, it is difficult to obtain sufficient research on humans, so the use of an animal model of IDD is very important to clarify the pathogenesis and treatment mechanism of the disease.
Methods: In this study, thirty 2-month-old mice were selected for operation to establish a coccygeal IDD model.
J Am Coll Surg
January 2025
Department of Surgery, Stanford University, Stanford, CA.
Background: Motion-tracking has been shown to correlate with expert and novice performance but has not been used for skill development. For skill development, performance goals must be defined. We hypothesize that using wearable sensor technology, motion tracking outcomes can be identified in those deemed practice-ready and used as benchmarks for precision learning.
View Article and Find Full Text PDFHeliyon
January 2025
BRITElab, Harry Perkins Institute of Medical Research, QEII Medical Centre Nedlands and Centre for Medical Research, The University of Western Australia, Perth, Australia.
Breast-conserving surgery accompanied by adjuvant radiotherapy is the standard of care for patients with early-stage breast cancer. However, re-excision is reported in 20-30 % of cases, largely because of close or involved tumor margins in the specimen. Several intraoperative tumor margin assessment techniques have been proposed to overcome this issue, however, none have been widely adopted.
View Article and Find Full Text PDFEnter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!